[起搏器患者的X线表现(作者译)]

[Roentgenographic findings in patients with pacemakers (author's transl)].

作者信息

du Mesnil de Rochemont W

出版信息

Rontgenblatter. 1976 Jan;29(1):5-12.

PMID:1251123
Abstract

Early roentgenographic documentation of the position of an implanted pacemaker by roentgenograms in two planes and fluoroscopy establishes the necessary basis for the roentgenographic examinations in identifying pacemaker malfunction. Early and late causes of pacemaker malfunction in 250 patients are described. Early complications arise at the tip of the electrode: displacement with or against the direction of blood flow, penetration or perforation of the myocardium and false implantation into the coronary sinus. Later causes of pacemaker failure have to be searched mainly in the electrode itself and the battery: damage to the cable or displacement of the battery. The problems of radiologic assessment of the state of charge of the battery are discussed.

摘要

早期通过两个平面的X线片和荧光透视对植入起搏器位置进行影像学记录,为通过X线检查识别起搏器故障奠定了必要基础。本文描述了250例患者起搏器故障的早期和晚期原因。早期并发症出现在电极尖端:电极随血流方向或逆血流方向移位、心肌穿透或穿孔以及误植入冠状窦。起搏器后期故障的原因主要在于电极本身和电池:电缆损坏或电池移位。文中还讨论了电池电量状态的放射学评估问题。