Kuzmiak Cherie M, Millnamow Gregory A, Qaqish Bahjat, Pisano Etta D, Cole Elodia B, Brown Marylee E
Department of Radiology, University of North Carolina, Box 7510, Chapel Hill, NC, 27514, USA.
Acad Radiol. 2002 Dec;9(12):1378-82. doi: 10.1016/s1076-6332(03)80664-5.
This study was performed to investigate whether full-field digital mammography (FFDM) is at least as accurate as screen-film mammography with respect to breast lesion characterization.
Seventy-nine breast surgical specimens were obtained by means of preoperative needle localization with surgical excision from 79 patients. The specimens were imaged with both screen-film mammography and FFDM. Six radiologists specialized in breast imaging analyzed both sets of images and characterized the visualized lesions on a five-point scale: 1, definitely not malignant; 2, probably not malignant; 3, possibly malignant; 4, probably malignant; and 5, definitely malignant. Receiver operating characteristic curve analysis of the data was then performed to assess for differences between modalities in the radiologists' ability to predict breast malignancy.
The areas under the receiver operating characteristic curves for the prediction of breast malignancy in surgical biopsy specimens were not statistically significantly different for FFDM and screen-film mammography.
The results demonstrate that with breast surgical specimens, FFDM is similar in diagnostic accuracy to screen-film mammography.
本研究旨在调查全视野数字乳腺摄影(FFDM)在乳腺病变特征描述方面是否至少与屏-片乳腺摄影一样准确。
通过术前针定位手术切除从79例患者获取79个乳腺手术标本。这些标本分别用屏-片乳腺摄影和FFDM进行成像。6名专门从事乳腺影像诊断的放射科医生分析了两组图像,并按照五点量表对可视化病变进行特征描述:1,肯定不是恶性;2,可能不是恶性;3,可能是恶性;4,可能是恶性;5,肯定是恶性。然后对数据进行受试者操作特征曲线分析,以评估两种模式在放射科医生预测乳腺恶性病变能力方面的差异。
FFDM和屏-片乳腺摄影在预测手术活检标本中乳腺恶性病变的受试者操作特征曲线下面积无统计学显著差异。
结果表明,对于乳腺手术标本,FFDM在诊断准确性方面与屏-片乳腺摄影相似。