A thermoluminescence system for the intercomparison of absorbed dose and radiation quality of X-rays with a HVL of 0-1 to 3-0 mm Cu.

作者信息

Puite K J

出版信息

Phys Med Biol. 1976 Mar;21(2):216-25. doi: 10.1088/0031-9155/21/2/003.

Abstract

A TLD system is described which permits the measurement of absorbed X-ray doses in a water phantom, together with the quality of the incident X-ray beam. The system has been developed for dose and energy intercomparison studies between centres with deep therapy X-ray machines. A distinction has been made between deep therapy treatments and treatments at or near the surface. For beam qualities with a HVL above 1-1 mm Cu, the deep therapy region, measurements have been performed at 5 cm depth in water using CaF2: Mn and LiF (TLD 700). For beam qualities with a HVL of 0-11-1-4 mm Cu, these qualities include the superficial therapy region, measurements with LiF dosemeters at 2 and 10 cm depth in water have been carried out. The uncertainty in the absorbed dose value for X-ray beams with a HVL of 0-11-3-0 mm Cu amounts to +/- 4%; the uncertainty in the determination of the effective energy of the incident beam is +/- 3 to +/- 7%.

摘要

描述了一种热释光剂量(TLD)系统,该系统可用于测量水模体中的X射线吸收剂量以及入射X射线束的品质。该系统是为使用深部治疗X射线机的各中心之间的剂量和能量比对研究而开发的。已对深部治疗和表面或接近表面的治疗进行了区分。对于半值层(HVL)高于1.1毫米铜的射束品质,即深部治疗区域,使用CaF2:Mn和LiF(TLD 700)在水中5厘米深度处进行了测量。对于HVL为0.11 - 1.4毫米铜的射束品质,这些品质包括浅部治疗区域,已使用LiF剂量计在水中2厘米和10厘米深度处进行了测量。对于HVL为0.11 - 3.0毫米铜的X射线束,吸收剂量值的不确定度为±4%;入射束有效能量测定的不确定度为±3%至±7%。