Comparison of the acidity of residual silanol groups in several liquid chromatography columns.

作者信息

Méndez Alberto, Bosch Elisabeth, Rosés Martí, Neue Uwe D

机构信息

Department de Química Analítica, Universitat de Barcelona, Diagonal 647, E-08028 Barcelona, Spain

出版信息

J Chromatogr A. 2003 Jan 31;986(1):33-44. doi: 10.1016/s0021-9673(02)01899-x.

Abstract

The silanol acidity of Waters Resolve C18, Waters Resolve silica, Waters Symmetry C18, Waters Symmetry silica, Waters XTerra MS C18 and underivatized XTerra columns has been measured from the retention of LiNO3 with a methanol/water (60:40) mobile phase buffered to different pH values. The Li+ cation is retained by cationic exchange with the background cation of the mobile phase (Na+) through the ionized silanols. The number of active silanols increases in the order: XTerra MS C18 << Symmetry C18 < underivatized XTerra << Resolve C18 < Resolve silica approximately equal to Symmetry silica. XTerra MS C18 does not present any residual silanol acidity up to s(s)pH 10.0 (pH in 60% methanol) as measured by LiNO3. The underivatized XTerra packing and Symmetry C18 present active silanols only at s(s)pH values higher than 7.0. For the other three columns, two different types of silanols with different acidity (s(s)pKa values about 3.5-4.6 and 6.2-6.8, respectively) have been observed. Symmetry C18 shows evidence of the presence of active basic sites that retain NO3- by anionic exchange.