Lázaro Elisabet, Ràfols Clara, Rosés Martí
Departament de Química Analítica, Universitat de Barcelona, Martí i Franquès 1, E-08028 Barcelona, Spain.
J Chromatogr A. 2008 Feb 29;1182(2):233-6. doi: 10.1016/j.chroma.2008.01.003. Epub 2008 Jan 8.
Residual silanol acidity and activity of one immobilized artificial membrane (IAM) column have been measured from the retention of LiNO(3) in the column with a methanol/buffer (1mM in Na(+)) (60:40, v/v) mobile phase buffered to different pH values. Just one type of silanol with pssK(a)=7.61 (near the pH limit recommended by the manufacturer) was found, although these silanols show large activity. The results obtained have been compared with those obtained previously for Resolve C18, Resolve Silica, Symmetry C18, Symmetry Silica, XTerra MS C18, underivatized XTerra, Lichrospher 100 RP-18, Purospher RP-18e, Luna C18 (2) and Chromolith Perfomance RP-18e, showing that the IAM column is similar to Luna C18 and Symmetry C18 in terms of silica quality, as measured by Li(+) retention. A warning about the use of IAM columns for emulation of biological systems at physiological pH 7 is given because the ionized silanols may contribute to the retention of some drugs at this pH.
通过在甲醇/缓冲液(Na⁺浓度为1mM)(体积比60:40)的流动相中,将硝酸锂保留在固定化人工膜(IAM)柱中,并将流动相缓冲至不同pH值,来测定该柱的残留硅醇酸度和活性。尽管这些硅醇表现出较大的活性,但仅发现一种pKa = 7.61的硅醇(接近制造商推荐的pH极限)。将所得结果与之前在Resolve C18、Resolve Silica、Symmetry C18、Symmetry Silica、XTerra MS C18、未衍生化的XTerra、Lichrospher 100 RP - 18、Purospher RP - 18e、Luna C18(2)和Chromolith Perfomance RP - 18e上获得的结果进行了比较,结果表明,就通过Li⁺保留测定的硅胶质量而言,IAM柱与Luna C18和Symmetry C18相似。由于离子化的硅醇可能会在生理pH 7时导致某些药物的保留,因此给出了关于在生理pH 7下使用IAM柱模拟生物系统的警告。