Geijer Håkan, Verdonck Bert, Beckman Karl-Wilhelm, Andersson Torbjörn, Persliden Jan
Department of Radiology, Orebro University Hospital, 701 85 Orebro, Sweden.
Eur Radiol. 2003 Mar;13(3):543-51. doi: 10.1007/s00330-002-1476-1. Epub 2002 Jun 14.
The aim of this study was optimization of the radiation dose-image quality relationship for a digital scanning method of scoliosis radiography. The examination is performed as a digital multi-image translation scan that is reconstructed to a single image in a workstation. Entrance dose was recorded with thermoluminescent dosimeters placed dorsally on an Alderson phantom. At the same time, kerma area product (KAP) values were recorded. A Monte Carlo calculation of effective dose was also made. Image quality was evaluated with a contrast-detail phantom and Visual Grading. The radiation dose was reduced by lowering the image intensifier entrance dose request, adjusting pulse frequency and scan speed, and by raising tube voltage. The calculated effective dose was reduced from 0.15 to 0.05 mSv with reduction of KAP from 1.07 to 0.25 Gy cm(2) and entrance dose from 0.90 to 0.21 mGy. The image quality was reduced with the Image Quality Figure going from 52 to 62 and a corresponding reduction in image quality as assessed with Visual Grading. The optimization resulted in a dose reduction to 31% of the original effective dose with an acceptable reduction in image quality considering the intended use of the images for angle measurements.
本研究的目的是优化脊柱侧弯X线摄影数字扫描方法的辐射剂量与图像质量的关系。该检查通过数字多图像平移扫描进行,并在工作站中重建为单个图像。使用置于Alderson体模背部的热释光剂量计记录入射剂量。同时,记录比释动能面积乘积(KAP)值。还进行了有效剂量的蒙特卡罗计算。使用对比细节体模和视觉分级评估图像质量。通过降低影像增强器入射剂量要求、调整脉冲频率和扫描速度以及提高管电压来降低辐射剂量。计算得到的有效剂量从0.15 mSv降至0.05 mSv,KAP从1.07 Gy cm²降至0.25 Gy cm²,入射剂量从0.90 mGy降至0.21 mGy。图像质量因子从52降至62,图像质量随之降低,视觉分级评估的图像质量也相应下降。考虑到图像用于角度测量的预期用途,优化后有效剂量降至原始剂量的31%,图像质量的降低在可接受范围内。