Gu W Y, Justiz M A
Tissue Biomechanics Lab, Department of Biomedical Engineering, College of Engineering, University of Miami, Coral Gables, FL 33124, USA.
J Biomech Eng. 2002 Dec;124(6):790-3. doi: 10.1115/1.1516571.
This paper describes a new apparatus and method for measuring swelling dependent electrical conductivity of charged hydrated soft tissues. The apparatus was calibrated using a conductivity standard. Swelling dependent specific conductivity of porcine annulus fibrosis (AF) samples was determined. The conductivity values for porcine AF were similar to those for human and bovine articular cartilage found in the literature. Results revealed a significant linear correlation between specific conductivity and water content for porcine AF tissues tested in phosphate buffered saline (PBS).
本文描述了一种用于测量带电水合软组织肿胀依赖性电导率的新装置和方法。该装置使用电导率标准进行校准。测定了猪纤维环(AF)样本的肿胀依赖性比电导率。猪AF的电导率值与文献中报道的人和牛关节软骨的电导率值相似。结果显示,在磷酸盐缓冲盐水(PBS)中测试的猪AF组织的比电导率与含水量之间存在显著的线性相关性。