Repeatability of ocular wavefront measurement.

作者信息

Davies Nigel, Diaz-Santana Luis, Lara-Saucedo David

机构信息

Department of Ophthalmology, The Middlesex Hospital, Mortimer Street, London, United Kingdom.

出版信息

Optom Vis Sci. 2003 Feb;80(2):142-50. doi: 10.1097/00006324-200302000-00010.

Abstract

PURPOSE

To assess the repeatability of measurements of ocular aberrations using wavefront sensing in a small group of observers and to assess the potential effect of measurement error on custom corneal correction due to this variability.

METHOD

A Shack-Hartmann wavefront sensor was used to measure the ocular wavefront in nine eyes. Head position was stabilized using a dental bite bar, and the pupil was centred using a cathode ray tube monitor and circular grating. Twenty Shack-Hartmann images were collected for each measurement. Each observer had three sets of measurements taken; the first and the second after careful alignment and the final after regrasping the bite bar in the same position as for the second measurement, but without pupil realignment. The modulation transfer functions for each set were calculated, and the effect of best-aligned custom treatments on the modulation transfer function was estimated.

RESULTS

There were highly statistically significant differences in a large number of Zernike modes between the three sets of measurements. The modulation transfer functions calculated for the residual wavefronts after aligned custom treatment were below the diffraction limit. The root mean square wavefront errors were consistently better for the residual wavefronts obtained using the realigned data than using data taken without pupil realignment.

CONCLUSIONS

Sequential measurement of ocular aberrations shows statistically significant differences in a large number of Zernike modes. If aberrations determined by a single measurement are to be used in a custom correction, the resulting modulation transfer function is likely to remain below the diffraction limit. Pupil realignment is critical in reduction of the residual root mean square wavefront values to a minimum.

摘要

目的

评估一小群观察者使用波前传感测量眼像差的可重复性,并评估由于这种变异性导致的测量误差对定制角膜矫正的潜在影响。

方法

使用夏克-哈特曼波前传感器测量9只眼睛的眼波前。使用牙托稳定头部位置,并使用阴极射线管显示器和圆形光栅使瞳孔居中。每次测量采集20张夏克-哈特曼图像。每位观察者进行三组测量;第一组和第二组在仔细对准后进行,最后一组在与第二次测量相同位置重新抓住牙托后进行,但不重新对准瞳孔。计算每组的调制传递函数,并估计最佳对准的定制治疗对调制传递函数的影响。

结果

三组测量之间在大量泽尼克模式上存在高度统计学显著差异。对准定制治疗后为残余波前计算的调制传递函数低于衍射极限。使用重新对准数据获得的残余波前的均方根波前误差始终优于未重新对准瞳孔时采集的数据。

结论

眼像差的顺序测量在大量泽尼克模式上显示出统计学显著差异。如果在定制矫正中使用单次测量确定的像差,则所得调制传递函数可能仍低于衍射极限。瞳孔重新对准对于将残余均方根波前值降至最低至关重要。