正常眼的泽尼克系数和均方根波前误差。

Normal-eye Zernike coefficients and root-mean-square wavefront errors.

作者信息

Salmon Thomas O, van de Pol Corina

机构信息

Northeastern State University, Tahlequah, Oklahoma, USA.

出版信息

J Cataract Refract Surg. 2006 Dec;32(12):2064-74. doi: 10.1016/j.jcrs.2006.07.022.

Abstract

PURPOSE

To compare aberrometry measurements from multiple sites and compute mean Zernike coefficients and root-mean-square (RMS) values for the entire data pool to serve as a reference set for normal, healthy adult eyes.

SETTING

Northeastern State University, Tahlequah, Oklahoma, USA.

METHODS

Data were collected from 10 laboratories that measured higher-order aberrations (HOAs) in normal, healthy adult eyes using Shack-Hartmann aberrometry (2560 eyes of 1433 subjects). Signed Zernike coefficients were scaled to pupil diameters of 6.0 mm, 5.0 mm, 4.0 mm, and 3.0 mm and corrected to a common wavelength of 550 nm. The mean signed and absolute Zernike coefficients across data sets were compared. Then, the following were computed: overall mean values for signed and absolute Zernike coefficients; polar Zernike magnitudes and RMS values for coma-like aberrations (Z(3)(+/-1) and Z(5)(+/-1) combined); spherical-like aberrations (Z(4)(0) and Z(6)(0) combined); and 3rd-, 4th-, 5th-, and 6th-order, and higher-order aberrations (orders 3 to 6).

RESULTS

The different data sets showed good agreement for Zernike coefficients values across most higher-order modes, with greater variability for Z(4)(0) and Z(3)(-1). The most prominent modes and their mean absolute values (6.0-mm pupil) were, respectively, Z(3)(-1) and 0.14 microm, Z(4)(0) and 0.13 microm, and Z(3)(-3) and 0.11 microm. The mean total higher-order RMS was 0.33 microm.

CONCLUSIONS

There was a general consensus for the magnitude of HOAs expected in normal adult human eyes. At least 90% of the sample had aberrations less than double the mean values reported here. These values can serve as a set of reference norms.

摘要

目的

比较多个测量点的像差测量结果,并计算整个数据集的平均泽尼克系数和均方根(RMS)值,作为正常健康成年人眼睛的参考数据集。

地点

美国俄克拉荷马州塔勒夸市东北州立大学。

方法

从10个实验室收集数据,这些实验室使用夏克-哈特曼像差仪测量正常健康成年人眼睛的高阶像差(HOAs)(1433名受试者的2560只眼睛)。有符号的泽尼克系数按比例缩放到6.0毫米、5.0毫米、4.0毫米和3.0毫米的瞳孔直径,并校正到550纳米的共同波长。比较各数据集的平均有符号和绝对泽尼克系数。然后,计算以下各项:有符号和绝对泽尼克系数的总体平均值;类慧差像差(Z(3)(+/-1)和Z(5)(+/-1)合并)、类球差像差(Z(4)(0)和Z(6)(0)合并)以及三阶、四阶、五阶和六阶及更高阶像差(3至6阶)的极坐标泽尼克幅值和RMS值。

结果

不同数据集在大多数高阶模式下的泽尼克系数值显示出良好的一致性,Z(4)(0)和Z(3)(-1)的变异性更大。最显著的模式及其平均绝对值(6.0毫米瞳孔)分别为Z(

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验