用于矫正高度近视的Artisan有晶状体眼人工晶状体植入术后的眼内光学质量。

Optical quality of the eye with the Artisan phakic lens for the correction of high myopia.

作者信息

Brunette Isabelle, Bueno Juan M, Harissi-Dagher Mona, Parent Mireille, Podtetenev Michel, Hamam Habib

机构信息

Department of Ophthalmology, University of Montreal, Montreal, Quebec, Canada.

出版信息

Optom Vis Sci. 2003 Feb;80(2):167-74. doi: 10.1097/00006324-200302000-00013.

Abstract

PURPOSE

To evaluate the optical quality of the eye before and after the insertion of an Artisan phakic intraocular lens for the treatment of high myopia.

METHODS

Consecutive patients implanted with the Artisan lens by a single surgeon between June 2001 and April 2002 were enrolled prospectively. One eye per subject was tested. The wavefront aberration was calculated from images recorded with a Hartmann-Shack sensor. This wavefront aberration was expressed as a Zernike polynomial expansion from the third up to the seventh order. Root mean square wavefront error was used as a parameter of optical quality. Point-spread function and modulation transfer function were also computed from the wavefront aberration.

RESULTS

The mean age of the four patients (four eyes) was 46 +/- 11 years. The preoperative mean spherical equivalent was -14.13 +/- 3.19 D (range, -20.50 to -9.75 D), with a best-corrected visual acuity of 20/25 or better in three of the four eyes. No complications were encountered. Postoperatively, the mean spherical equivalent was -0.22 +/- 0.30 D (range, -0.75 to +0.38 D). An uncorrected visual acuity of 20/40 or better was observed in three eyes. Overall, for each combination of order (third, fourth, and fifth to seventh) and pupil size (3, 4, and 5 mm), the mean postoperative root mean square values for the four subjects were lower than the mean preoperative values. However, because of the small size of the study population (four patients), this improvement did not reach a statistically significant level.

CONCLUSIONS

Preliminary data using the Hartmann-Shack wavefront sensor have not revealed a tendency toward deterioration of the optical performance after the insertion of an Artisan lens for the treatment of high myopia. The Hartmann-Shack sensor was a useful tool for the objective assessment of the image optical quality of eyes with a phakic intraocular lens.

摘要

目的

评估植入Artisan有晶状体眼人工晶状体治疗高度近视前后眼睛的光学质量。

方法

前瞻性纳入2001年6月至2002年4月间由同一位外科医生植入Artisan晶状体的连续患者。每位受试者测试一只眼睛。通过Hartmann-Shack传感器记录的图像计算波前像差。该波前像差用从三阶到七阶的Zernike多项式展开表示。均方根波前误差用作光学质量参数。还从波前像差计算点扩散函数和调制传递函数。

结果

4例患者(4只眼)的平均年龄为46±11岁。术前平均等效球镜度为-14.13±3.19 D(范围为-20.50至-9.75 D),4只眼中有3只眼的最佳矫正视力为20/25或更好。未发生并发症。术后,平均等效球镜度为-0.22±0.30 D(范围为-0.75至+0.38 D)。3只眼的裸眼视力为20/40或更好。总体而言,对于阶数(三阶、四阶以及五阶至七阶)和瞳孔大小(3、4和5 mm)的每种组合,4名受试者术后的均方根平均值均低于术前平均值。然而,由于研究人群规模较小(4例患者),这种改善未达到统计学显著水平。

结论

使用Hartmann-Shack波前传感器的初步数据未显示植入Artisan晶状体治疗高度近视后光学性能有恶化趋势。Hartmann-Shack传感器是客观评估有晶状体眼人工晶状体眼图像光学质量的有用工具。