Mitchell G A, Pochron S P, Smutny P V, Guity R
Clin Chem. 1976 May;22(5):647-9.
Radoiassay values for folate in protein-denautured (extracted) sera are compared with values obtained by individual serum blank correction. Lower folate values are reported for sera extracted at pH 1.3, 7.6, 10.5, and 12.4 when pterolyglutamic acid (PGA) calibration standards are used. The results are unchanged when ascorbate added to the sera is increased from 1 to 20 g/liter and when [3H]PGA is substituted for 125I-labeled PGA. However, a higher mean value for extracted serum is reported when N-5-methyltetrahydrofolic acid calibration standards are substituted for PGA. Because both patients' samples and assay calibrators were subjected to the same extraction methods, our results suggest that serum folate behaves differently from PGA during the extractions.
将蛋白质变性(提取)血清中叶酸的放射免疫分析值与通过个体血清空白校正获得的值进行比较。当使用蝶酰谷氨酸(PGA)校准标准品时,报告了在pH 1.3、7.6、10.5和12.4下提取的血清中叶酸值较低。当血清中添加的抗坏血酸从1 g/升增加到20 g/升以及用[3H]PGA替代125I标记的PGA时,结果不变。然而,当用N-5-甲基四氢叶酸校准标准品替代PGA时,报告的提取血清平均值较高。由于患者样本和分析校准品都采用相同的提取方法,我们的结果表明,在提取过程中血清叶酸的行为与PGA不同。