Woolfson Peter I, Pullan Brian R, Lewis Philip S
Department of Medicine, Stepping Hill Hospital, Hazel Grove, Stockport, UK.
Biomed Instrum Technol. 2003 Jan-Feb;37(1):41-6. doi: 10.2345/0899-8205(2003)37[41:BFMFPP]2.0.CO;2.
An air plethysmograph with a sensitive phototransducer was constructed so that plethysmographic volume-change pulsations could be displayed in detail without using venous occlusion. Software was developed to allow analysis of the pulses using a modification of the backward extrapolation technique. This allowed calculation of the forward arterial blood flow and noninvasive derivation of the resting arterial flow waveform. There is good reproducibility of the technique, with 8% variability between pairs of measurements at rest and 4% variability after hand exercise. Direct comparison made with blood flows measured by venous occlusion plethysmography showed good average agreement. The mean blood flow for venous occlusion (rest and exercise) was 0.76 +/- 0.07 mL/beat (mean +/- SEM), and the mean blood flow for backward extrapolation (rest and exercise) was 0.74 +/- 0.09 mL/beat (mean +/- SEM). This corresponds to 3.86 +/- 0.36 mL/min/100 mL and 3.76 +/- 0.46 mL/min/100 mL, respectively. Important assumptions when using this method are that venous return is constant and that forward arterial flow is over before the end of the cardiac cycle.
构建了一种带有灵敏光电换能器的空气体积描记器,这样无需使用静脉阻断就能详细显示体积描记的容积变化搏动。开发了软件,以便使用改进的反向外推技术对搏动进行分析。这使得能够计算正向动脉血流,并无创推导静息动脉血流波形。该技术具有良好的可重复性,静息时成对测量之间的变异性为8%,手部运动后为4%。与静脉阻断体积描记法测量的血流进行直接比较,显示出良好的平均一致性。静脉阻断(静息和运动)的平均血流为0.76±0.07毫升/搏(平均值±标准误),反向外推(静息和运动)的平均血流为0.74±0.09毫升/搏(平均值±标准误)。这分别对应于3.86±0.36毫升/分钟/100毫升和3.76±0.46毫升/分钟/100毫升。使用该方法时的重要假设是静脉回流恒定,且正向动脉血流在心动周期结束前结束。