A HPLC method was developed for the simultaneous determination of organic peroxides and hydrogen peroxide with peroxyoxalate chemiluminescence (PO-CL) detection following on-line UV irradiation. Organic peroxides [i.e., benzoyl peroxide (BP), tert.-butyl hydroperoxide (BHP), tert.-butyl perbenzoate (BPB), cumene hydroperoxide (CHP)] were UV irradiated (254 nm, 15 W) to generate hydrogen peroxide, which was determined by PO-CL detection. The conditions for UV irradiation and PO-CL detection were optimized by a flow injection analysis (FIA) system. Generation of hydrogen peroxide from peroxides with on-line UV irradiation also was confirmed by the FIA system by incorporating an enzyme column reactor immobilized with catalase. The separation of four organic peroxides and hydrogen peroxide by HPLC was accomplished isocratically on an ODS column within 30 min. The detection limits (signal-to-noise ratio=3) were 1.1 microM for hydrogen peroxide, 6.8 microM for BP, 31.3 microM for BHP, 7.5 microM for BPB and 1.3 microM for CHP. The proposed method was applied to the determination of BP in wheat flour.
建立了一种高效液相色谱法,用于在线紫外照射后通过过氧草酸酯化学发光(PO-CL)检测同时测定有机过氧化物和过氧化氢。有机过氧化物[即过氧化苯甲酰(BP)、叔丁基过氧化氢(BHP)、叔丁基过苯甲酸酯(BPB)、异丙苯过氧化氢(CHP)]经紫外照射(254 nm,15 W)生成过氧化氢,通过PO-CL检测进行测定。通过流动注射分析(FIA)系统对紫外照射和PO-CL检测的条件进行了优化。通过结合固定有过氧化氢酶的酶柱反应器,FIA系统也证实了过氧化物经在线紫外照射生成过氧化氢。在ODS柱上通过等度洗脱在30分钟内完成了四种有机过氧化物和过氧化氢的高效液相色谱分离。检测限(信噪比 = 3):过氧化氢为1.1 μM,BP为6.8 μM,BHP为31.3 μM,BPB为7.5 μM,CHP为1.3 μM。所提出的方法应用于小麦粉中BP的测定。