Anderson S T, Davis B B
J Thorac Cardiovasc Surg. 1976 May;71(5):698-703.
One hundred and seventy-three unipolar pulse generator systems were tested in 169 patients to determine the incidence of sensing of muscle potentials. One hundred and twenty-two implants in the pectoral region were suitable for assessment and 60 (49 per cent) demonstrated sensing. Ventricular-inhibited units sensed muscle potentials in 69 per cent (47 of 78), with 9 patients complaining of related symptoms. Ventricular-synchronous units sensed in 30 per cent (13 of 34) with no patient manifesting symptoms.
对169例患者的173个单极脉冲发生器系统进行了测试,以确定肌肉电位感知的发生率。胸肌区域的122个植入物适合进行评估,其中60个(49%)显示有感知。心室抑制型装置中,69%(78个中的47个)感知到肌肉电位,有9例患者抱怨有相关症状。心室同步型装置中,30%(34个中的13个)有感知,无患者出现症状。