Völk Marcus, Angele Peter, Hamer Okka, Feuerbach Stefan, Strotzer Michael
Department of Diagnostic Radiology. University Hospital of Regensburg, Germany.
Invest Radiol. 2003 Apr;38(4):189-92. doi: 10.1097/01.RLI.0000057028.63568.85.
The purpose of this study was to evaluate the image quality of composed long-leg examinations with a large-area, flat-panel x-ray detector.
Thirty-five consecutive patients were included in this study. All images were obtained with a kilovoltage setting identical with conventional radiographies of speed class 400; amperage values were reduced by 50% compared with standard dose. After acquisition, the images were transferred to a workstation where the whole image was reconstructed using a generalized correlation method. Images were presented to 3 observers. Examination quality was ranked on a 3-point scale: 1 = no manual adjustment necessary; 2 = composition required manual correction; 3 = no composition possible.
Patient rankings were 31/35 (88.6%) in category 1, 3/35 (8.6%) in category 2, and 1/35 (2.8%) in category 3 (primarily due to an application error).
The analysis of the first clinical examinations of long-leg radiographies with a 43 cm x 43 cm flat-panel detector demonstrates very good reliability of the digital image composition.
本研究旨在评估使用大面积平板X射线探测器进行的组合式长腿检查的图像质量。
本研究纳入了35例连续患者。所有图像均在与400速级常规X线摄影相同的千伏设置下获得;与标准剂量相比,电流值降低了50%。采集后,图像被传输到工作站,在那里使用广义相关方法重建整个图像。图像展示给3名观察者。检查质量按3分制进行排名:1 = 无需手动调整;2 = 构图需要手动校正;3 = 无法构图。
患者排名中,1类占31/35(88.6%),2类占3/35(8.6%),3类占1/35(2.8%)(主要由于应用错误)。
对使用43 cm×43 cm平板探测器进行的长腿X线摄影的首次临床检查分析表明,数字图像构图具有非常好的可靠性。