Zwanenburg Jaco J M, Kuijer Joost P A, Marcus J Tim, Heethaar Robert M
Department of Physics & Medical Technology, VU University Medical Center, PO Box 7057, 1007 MB Amsterdam, The Netherlands.
Magn Reson Med. 2003 Apr;49(4):722-30. doi: 10.1002/mrm.10422.
A method is presented that combines steady-state free precession (SSFP) cine imaging with myocardial tagging. Before the tagging preparation at each ECG-R wave, the steady-state magnetization is stored as longitudinal magnetization by an alpha/2 flip-back pulse. Imaging is continued immediately after tagging preparation, using linearly increasing startup angles (LISA) with a rampup over 10 pulses. Interleaved segmented k-space ordering is used to prevent artifacts from the increasing signal during the LISA rampup. First, this LISA-SSFP method was evaluated regarding ghost artifacts from the steady-state interruption by comparing LISA with an alpha/2 startup method. Next, LISA-SSFP was compared with spoiled gradient echo (SGRE) imaging, regarding tag contrast-to-noise ratio and tag persistence. The measurements were performed in phantoms and in six subjects applying breathhold cine imaging with tagging (temporal resolution 51 ms). The results show that ghost artifacts are negligible for the LISA method. Compared to the SGRE reference, LISA-SSFP was two times faster, with a slightly better tag contrast-to-noise. Additionally, the tags persisted 126 ms longer with LISA-SSFP than with SGRE imaging. The high efficiency of LISA-SSFP enables the acquisition of complementary tagged (CSPAMM) images in a single breathhold.
本文介绍了一种将稳态自由进动(SSFP)电影成像与心肌标记相结合的方法。在每次心电图R波进行标记准备之前,通过一个α/2反转脉冲将稳态磁化强度存储为纵向磁化强度。在标记准备完成后立即继续成像,使用线性增加的起始角度(LISA),在10个脉冲内进行斜坡上升。采用交错分割的k空间排序来防止在LISA斜坡上升期间由于信号增加而产生伪影。首先,通过将LISA与α/2起始方法进行比较,评估了这种LISA-SSFP方法关于稳态中断产生的鬼影伪影。接下来,在标记对比度噪声比和标记持久性方面,将LISA-SSFP与扰相梯度回波(SGRE)成像进行了比较。测量在体模和六名受试者中进行,采用屏气电影成像并进行标记(时间分辨率51毫秒)。结果表明,LISA方法产生的鬼影伪影可忽略不计。与SGRE参考相比,LISA-SSFP快两倍,标记对比度噪声略好。此外,LISA-SSFP的标记持续时间比SGRE成像长126毫秒。LISA-SSFP的高效性使得能够在单次屏气中采集互补标记(CSPAMM)图像。