Kellman Peter, Guttman Michael A, Herzka Daniel A, McVeigh Elliot R
Laboratory of Cardiac Energetics, National Institutes of Health, National Heart, Lung and Blood Institute, Bethesda, Maryland 20892-1061, USA.
Magn Reson Med. 2002 Dec;48(6):1076-80. doi: 10.1002/mrm.10309.
Steady-state free precession (SSFP) has recently proven to be valuable for cardiac imaging due to its high signal-to-noise ratio and blood-myocardium contrast. Data acquired using ECG-triggered, segmented sequences during the approach to steady-state, or return to steady-state after interruption, may have ghost artifacts due to periodic k-space distortion. Schemes involving several preparatory RF pulses have been proposed to restore steady-state, but these consume imaging time during early systole. Alternatively, the phased-array ghost elimination (PAGE) method may be used to remove ghost artifacts from the first several frames. PAGE was demonstrated for cardiac cine SSFP imaging with interrupted steady-state using a simple alpha/2 magnetization preparation and storage scheme and a spatial tagging preparation.
稳态自由进动(SSFP)由于其高信噪比和血液-心肌对比度,最近已被证明对心脏成像很有价值。在接近稳态或中断后恢复稳态期间,使用心电图触发的分段序列采集的数据可能会因周期性k空间失真而产生鬼影伪影。已经提出了涉及几个预备射频脉冲的方案来恢复稳态,但这些方案在早期收缩期会消耗成像时间。或者,可以使用相控阵鬼影消除(PAGE)方法从最初的几帧中去除鬼影伪影。使用简单的α/2磁化准备和存储方案以及空间标记准备,对中断稳态的心脏电影SSFP成像演示了PAGE。