Peters Christopher L, Craig Marcis A, Mohr R Alexander, Bachus Kent N
Department of Orthopedics, University of Utah School of Medicine, Salt Lake City, UT 84112, USA.
Clin Orthop Relat Res. 2003 Apr(409):158-68. doi: 10.1097/01.blo.0000058638.94987.20.
Despite excellent outcomes with cemented tibial components in total knee arthroplasty, it still is debated whether the tibial stem should be cemented and what the optimal tibial stem design should be. Proponents of full cementation of the tibial stem and component state that better short-term and long-term component fixation is achieved when full cementation is used. Advocates for surface cementation contend that sufficient implant stability is achieved without the increased bone loss that occurs at revision and the stress shielding thought to be linked with cemented stems. This biomechanical cadaver study compared initial fixation and cement penetration depth in fully cemented versus surface cemented tibial trays with two different stem geometries (cruciate and I-beam) and compared two stem designs (cruciate and I-beam) fixed with surface cementation. Under an eccentric load, simulating three times body weight for 6000 cycles, there seems to be no difference in the micromotion of either tibial component implanted with surface or full cementation. Additionally, no difference in the average depth of cement penetration was detected between fixation techniques or stem types. The initial fixation stability of the surface cement technique seems correlated to the depth of cement penetration into proximal tibial surface. The current data support other studies which indicate that stability of surface-cemented tibial components may be related to the depth of cement penetration.
尽管在全膝关节置换术中使用骨水泥固定的胫骨组件取得了优异的效果,但对于胫骨柄是否应使用骨水泥固定以及最佳的胫骨柄设计仍存在争议。支持胫骨柄和组件完全骨水泥固定的人认为,采用完全骨水泥固定可实现更好的短期和长期组件固定。支持表面骨水泥固定的人则认为,在翻修时不会出现骨量增加丢失以及应力遮挡(认为这与骨水泥柄有关)的情况下,可实现足够的植入物稳定性。这项生物力学尸体研究比较了采用两种不同柄几何形状(十字形和工字梁形)的完全骨水泥固定与表面骨水泥固定的胫骨托的初始固定情况和骨水泥渗透深度,并比较了采用表面骨水泥固定的两种柄设计(十字形和工字梁形)。在模拟三倍体重的偏心载荷下加载6000次循环,无论是采用表面骨水泥固定还是完全骨水泥固定植入的胫骨组件,其微动情况似乎没有差异。此外,在固定技术或柄类型之间未检测到骨水泥渗透平均深度的差异。表面骨水泥技术的初始固定稳定性似乎与骨水泥向胫骨近端表面的渗透深度相关。目前的数据支持其他研究,这些研究表明表面骨水泥固定的胫骨组件的稳定性可能与骨水泥渗透深度有关。