Masimo和Nellcor脉搏血氧饱和度传感器在婴儿护理中的使用寿命。

Longevity of Masimo and Nellcor pulse oximeter sensors in the care of infants.

作者信息

Erler Thomas, Avenarius Stefan, Wischniewski Esther, Schmidt Katerina, Kläber Hans-Georg

机构信息

Department of Pediatrics, Carl-Thiem-Klinikum, Cottbus, Germany.

出版信息

J Perinatol. 2003 Mar;23(2):133-5. doi: 10.1038/sj.jp.7210874.

Abstract

OBJECTIVE

Pulse oximetry is a standard of care for monitoring oxygenation in neonates. Associated with the use of pulse oximetry is the cost of patient sensors, especially if the sensor is designed for single-patient use. Pulse oximetry monitoring of sick newborns is routine and often lengthy and, if the pulse oximeter sensor is short-lived, can result in a significant portion in the cost of intensive care.

METHODS

We evaluated, in the NICUs of two hospitals and one step-down nursery, the useful life of disposable neonatal pulse oximeter sensors from two manufacturers: Masimo and Nellcor. The only requisites were ethics committee approval and need for monitoring. The timed of PO sensor placement and replacement were noted along with the reason for changing the sensor. The standard care practices for PO and sensor use in the respective institutions were followed.

RESULTS

A total of 835.5 patient days of monitoring were accumulated with 65 infants in the Masimo group and 56 using Nellcor. The Masimo Neo sensors had over twice (2.33) the useful life of the Nellcor N-25 (9.05+/-4.4 versus 3.9+/-2.3 days (range of 7.2-11.8 and 2.5-5.8 days, respectively, p<0.05)). The magnitude of useful life between the two institutions was not significantly different in the Masimo group (2.35- versus 2.22-fold). PO sensors were replaced due to impaired adhesion (38 Masimo and 32 Nellcor) and no signal (six Masimo and four Nellcor).

CONCLUSIONS

We found a more than two-fold increase in the life of Masimo versus Nellcor sensors. This difference was consistent between various caregivers in multiple settings and corroborates the experience of another, more limited study. A cost savings should result from the use of Masimo versus Nellcor disposable pulse oximeter sensors in neonatal routine care

摘要

目的

脉搏血氧饱和度测定法是监测新生儿氧合作用的护理标准。与脉搏血氧饱和度测定法的使用相关的是患者传感器的成本,特别是如果该传感器专为单人使用而设计。对患病新生儿进行脉搏血氧饱和度监测是常规操作,而且通常持续时间较长,如果脉搏血氧仪传感器使用寿命短,可能会导致重症监护成本的很大一部分。

方法

我们在两家医院的新生儿重症监护病房(NICU)和一个过渡性保育室中,评估了来自两家制造商(Masimo和Nellcor)的一次性新生儿脉搏血氧仪传感器的使用寿命。唯一的要求是获得伦理委员会的批准以及有监测需求。记录脉搏血氧饱和度(PO)传感器放置和更换的时间以及更换传感器的原因。遵循各机构中PO和传感器使用的标准护理规范。

结果

Masimo组有65名婴儿,Nellcor组有56名婴儿,总共积累了835.5个患者监测日。Masimo Neo传感器的使用寿命是Nellcor N-25的两倍多(2.33倍)(分别为9.05±4.4天和3.9±2.3天,范围分别为7.2 - 11.8天和2.5 - 5.8天,p<0.05)。在Masimo组中,两个机构之间使用寿命的差异不显著(2.35倍与2.22倍)。更换PO传感器的原因是粘贴受损(Masimo组38例,Nellcor组32例)和无信号(Masimo组6例,Nellcor组4例)。

结论

我们发现Masimo传感器的使用寿命比Nellcor传感器增加了两倍多。这种差异在多个环境中的不同护理人员之间是一致的,并且证实了另一项更有限研究的经验。在新生儿常规护理中使用Masimo一次性脉搏血氧仪传感器与Nellcor相比应能节省成本