Nitu-Whalley I C, Lee C A, Brown S A, Riddell A, Hermans C
Katharine Dormandy Haemophilia Centre and Haemostasis Unit, Royal Free University College London, Pond Street, London, UK.
Haemophilia. 2003 May;9(3):298-302. doi: 10.1046/j.1365-2516.2003.00755.x.
Determination of the closure time (CT) with the platelet function analyser (PFA-100) is a useful screening test for von Willebrand's disease (VWD) but its role in the characterization of VWD is not well established. We studied the relationship between the prolongation of the CT with adenosine diphosphate (ADP) (CT-ADP) and epinephrine (CT-EPI) cartridges and the von Willebrand factor (VWF) in 53 patients with VWD. We found that a relatively small percentage of the prolongation of the CT-ADR and CT-ADP (16 and 29%, respectively) was determined by a reduction in VWF levels. The CT-ADP was significantly more prolonged in the presence of qualitative defects in VWF but could not discriminate between the VWD subtypes. The ABO blood group had no effect on the prolongation of the CT or the bleeding time. In conclusion, the PFA-100 appears of little use in the characterization of severity and subtype of VWD.
使用血小板功能分析仪(PFA - 100)测定封闭时间(CT)是血管性血友病(VWD)的一项有用的筛查试验,但其在VWD特征描述中的作用尚未完全明确。我们研究了53例VWD患者中,使用二磷酸腺苷(ADP)(CT - ADP)和肾上腺素(CT - EPI)检测卡时CT的延长与血管性血友病因子(VWF)之间的关系。我们发现,CT - ADR和CT - ADP延长的比例相对较小(分别为16%和29%)是由VWF水平降低所致。在VWF存在定性缺陷时,CT - ADP显著延长,但无法区分VWD的亚型。ABO血型对CT的延长或出血时间没有影响。总之,PFA - 100在VWD严重程度和亚型的特征描述中似乎用处不大。