Suppr超能文献

计算机记录与验证系统对放射治疗错误的促成作用。

Facilitation of radiotherapeutic error by computerized record and verify systems.

作者信息

Patton Gregory A, Gaffney David K, Moeller John H

机构信息

Northwest Cancer Specialists, PC, Portland, OR 97227, USA.

出版信息

Int J Radiat Oncol Biol Phys. 2003 May 1;56(1):50-7. doi: 10.1016/s0360-3016(02)04418-8.

Abstract

PURPOSE

To examine the impact of computerized record and verify (R&V) systems on types of radiotherapeutic error.

MATERIALS AND METHODS

Radiation therapy treatment errors reported by therapists at the University of Utah between July 1, 1999 and June 30, 2000 were retrospectively reviewed.

RESULTS

During a 1-year period in which 22,542 external beam radiation therapy treatments were administered, 38 treatment errors (representing 0.17% of external beam treatments administered during this period) were identified and reviewed. Nine cases (0.04% of treatments) representing four types of record and verify (R&V)-related errors were identified, in which the department's R&V system played a contributory role in the treatment error.

CONCLUSIONS

The common denominator among these R&V-related errors was excessive reliance upon the computer system by therapists. R&V systems eliminate some, but not all, pathways of radiotherapeutic error. Although R&V systems have assumed a crucial role in the precise and reproducible delivery of increasingly complex radiation therapy treatments, their inability to eradicate all radiotherapeutic errors coupled with their parallel ability to facilitate certain mistakes mandates vigilance on the part of the radiation therapy team. Radiation therapy treatment procedures must preserve careful oversight of R&V functions to minimize prospects for treatment error.

摘要

目的

研究计算机记录与验证(R&V)系统对放射治疗错误类型的影响。

材料与方法

回顾性分析了犹他大学治疗师在1999年7月1日至2000年6月30日期间报告的放射治疗失误。

结果

在为期1年的时间里,共进行了22542次外照射放射治疗,发现并审查了38次治疗失误(占该期间外照射治疗的0.17%)。其中9例(占治疗的0.04%)代表4种与记录和验证(R&V)相关的失误,科室的R&V系统在这些治疗失误中起到了促成作用。

结论

这些与R&V相关的失误的共同特征是治疗师过度依赖计算机系统。R&V系统消除了部分但并非所有的放射治疗失误途径。尽管R&V系统在越来越复杂的放射治疗的精确和可重复实施中发挥了关键作用,但它们无法根除所有放射治疗失误,且同时具有助长某些错误的能力,这就要求放射治疗团队保持警惕。放射治疗程序必须对R&V功能进行仔细监督,以尽量减少治疗失误的可能性。