Cuddihy E F, Moacanin J, Roschke E J, Harrison E C
J Biomed Mater Res. 1976 May;10(3):471-81. doi: 10.1002/jbm.820100314.
Dynamic shear modulus G' was measured throughout the volume of three nonvariant silicone rubber poppets which were recovered from aortic prosthetic heart valves that had been implanted for 4 days, 52 days, and 8 years. Similar measurements were obtained for two unused silicone rubber poppets. Although the recovered poppets exhibited no obvious physical evidence of damage, the silicone rubber had undergone in vivo degradation throughout the poppet volume as indicated by decreases in modulus. The measurements also indicate that the poppet surface degrades at a rate faster than the core. Further, comparison with data reported in the literature suggests that the surfaces of variant poppets degrade at a rate faster than the surfaces of nonvariant poppets.
对从已植入4天、52天和8年的主动脉人工心脏瓣膜中回收的三个不变形硅橡胶瓣叶的整个体积进行了动态剪切模量G'的测量。对两个未使用的硅橡胶瓣叶也进行了类似的测量。尽管回收的瓣叶没有明显的物理损伤迹象,但如模量降低所示,硅橡胶在瓣叶整个体积内都经历了体内降解。测量还表明,瓣叶表面的降解速度比核心快。此外,与文献报道的数据比较表明,变形瓣叶的表面降解速度比不变形瓣叶的表面快。