Iplikçioğlu Haldun, Akça Kivanç, Cehreli Murat C, Sahin Salme
Department of Prosthodontics, Faculty of Dentistry, Hacettepe University, Sihhiye, Ankara, Turkey.
Int J Oral Maxillofac Implants. 2003 Mar-Apr;18(2):258-65.
To understand the mechanical and biomechanical behavior of dental Implants, validation of stress and strain measurements is required. The objective of this study was to compare a non-linear finite element stress analysis with in vitro strain gauge measurements on strains in an implant-abutment complex.
Strain gauges were bonded to an Implant-abutment complex and embedded in polymethylmethacrylate resin. A force of 75 N was applied vertically and laterally in separate load cases, and strains were recorded with a strain indicator. Then, a finite element model of the strain gauge model was constructed. Contact analysis with normal contact detection and separation behavior was performed between the Implant and the abutment. The same loading protocol was followed, and strains were recorded at regions where gauges were bonded.
Under vertical loading, the qualification and quantification of strains were similar in both methods. Under lateral loading, the measurement of strains on the abutment and in the resin were similar in both methods. However, strains on the implant collar as measured by non-linear finite element analysis were higher.
There is a compatibility between non-linear finite element stress analysis and in vitro strain gauge analysis on the measurement of strains under vertical loading. However, there are differences between the methods in the quantification of strains on the collar of Implants under lateral loading.
为了解牙种植体的力学和生物力学行为,需要对应力和应变测量进行验证。本研究的目的是比较种植体 - 基台复合体中应变的非线性有限元应力分析与体外应变片测量结果。
将应变片粘贴在种植体 - 基台复合体上,并嵌入聚甲基丙烯酸甲酯树脂中。在不同加载情况下分别垂直和侧向施加75 N的力,并用应变指示器记录应变。然后,构建应变片模型的有限元模型。在种植体和基台之间进行具有法向接触检测和分离行为的接触分析。遵循相同的加载方案,并在粘贴应变片的区域记录应变。
在垂直加载下,两种方法中应变的定性和定量相似。在侧向加载下,两种方法中基台和树脂中应变的测量相似。然而,通过非线性有限元分析测量的种植体颈部应变较高。
在垂直加载下测量应变时,非线性有限元应力分析与体外应变片分析之间具有兼容性。然而,在侧向加载下,两种方法在种植体颈部应变的定量方面存在差异。