Haeckel R
J Clin Chem Clin Biochem. 1976 Apr;14(4):165-71. doi: 10.1515/cclm.1976.14.1-12.165.
The enzymatic determination of the uric acie concentration in urine and serum according to Kageyama ((1971), Clin. Chim. Acta 31, 421--426), which excludes the deproteinization of samples, was adapted to the C4 automatic analyzer (Perkin-Elmer). The reliability of this procedure and its correlation with an UV-method were investigated. The interaction from-sample-to-sample was considerable, but this could be reduced by the addition of Brij-35. The recovery of uric acid added to protein-containing samples (about 96%) was better than with the UV-method. Novaminsulfone was the only substance tested which interfered significantly.