Umeda Naoya, Saito Masanobu, Sugano Nobuhiko, Ohzono Kenji, Nishii Takashi, Sakai Takashi, Yoshikawa Hideki, Ikeda Daisaku, Murakami Atsushi
Department of Orthopaedic Surgery, Hoshigaoka Koseinenkin Hospital, 4-8-1 Hoshigaoka, Hirakata 573-8511, Japan.
J Orthop Sci. 2003;8(3):381-6. doi: 10.1007/s10776-002-0635-2.
Stress analysis on the femur after insertion of a femoral prosthesis with a modular interchangeable neck with adjustable neck version was performed using in vitro experimental procedures and cemented strain gauges. Although strain on the anterior and posterior aspects was less than on the medial and lateral aspects, it was more markedly affected by adjusting the interchangeable neck from straight to anteverted (or retroverted) than strain on the medial and lateral aspects, particularly around the stem tip. There was a highly significant correlation between compressive strain on the side toward which the prosthetic neck was oriented and the extent of neck version, whether anteverted or retroverted. Insertion of a femoral implant in excessive anteversion (or retroversion) may cause noticeably increased stress on the anterior and posterior aspects of the proximal femoral shaft.
使用体外实验程序和粘贴式应变片,对带有可调节颈段版本的模块化可互换颈段的股骨假体植入后股骨进行应力分析。尽管前后侧的应变小于内外侧的应变,但与内外侧的应变相比,将可互换颈段从直型调整为前倾角(或后倾角)时,其应变受影响更显著,尤其是在柄尖周围。假体颈段所朝向一侧的压缩应变与颈段版本程度(无论是前倾角还是后倾角)之间存在高度显著的相关性。股骨植入物过度前倾角(或后倾角)插入可能会导致股骨近端干骺端前后侧应力明显增加。