Wu Raymond K, McGinley Patton H
Department of Radiation Oncology & Biophysics, Eastern Virginia Medical School, 600 Gresham Drive, Norfolk, Virginia 23507, USA.
J Appl Clin Med Phys. 2003 Spring;4(2):162-71. doi: 10.1120/jacmp.v4i2.2532.
Neutron and photon dose equivalents at various points along the mazes of the vaults for two 15 MV linear accelerators were measured. The measurements were made with the machines set at various gantry angles with and without the scattering phantom, and with the collimators set at the maximum and the minimum field sizes. Neutron dose equivalent measurements were made for five other accelerator vaults. Empirical equations were used to fit the dose data at points along the center of the maze, at 1 m above floor level, with the primary radiation beam pointing downward. It is reported here that both the capture gamma and the neutron dose attenuations along the maze are in agreement with the literature. The neutron dose is dependent on the square root of the ratio of the cross-sectional areas of the inner maze entrance and the maze. The tenth value distance (T(N)) is proportional to the square root of the cross-sectional area of the maze.
测量了两台15兆伏直线加速器防护室迷宫不同位置处的中子和光子剂量当量。测量时,机器设置了不同的机架角度,有散射体模和无散射体模,准直器设置为最大和最小射野尺寸。还对另外五个加速器防护室进行了中子剂量当量测量。利用经验公式拟合了主辐射束向下时,迷宫中心线上距地面1米处各点的剂量数据。本文报道,沿迷宫的俘获γ剂量和中子剂量衰减与文献一致。中子剂量取决于内迷宫入口与迷宫横截面积之比的平方根。十分之一值距离(T(N))与迷宫横截面积的平方根成正比。