Falcão R C, Facure A, Silva A X
Comissão Nacional de Energia Nuclear, Rua Gal. Severiano 90, sala 405, 22294-900 Rio de Janeiro, RJ, Brazil.
Radiat Prot Dosimetry. 2007;123(3):283-7. doi: 10.1093/rpd/ncl144. Epub 2006 Sep 27.
Currently, teletherapy machines of cobalt and caesium are being replaced by linear accelerators. The maximum photon energy in these machines can vary from 4 to 25 MeV, and one of the great advantages of these equipments is that they do not have a radioactive source incorporated. High-energy (E > 10 MV) medical linear accelerators offer several physical advantages over lower energy ones: the skin dose is lower, the beam is more penetrating, and the scattered dose to tissues outside the target volume is smaller. Nevertheless, the contamination of undesirable neutrons in the therapeutic beam, generated by the high-energy photons, has become an additional problem as long as patient protection and occupational doses are concerned. The treatment room walls are shielded to attenuate the primary and secondary X-ray fluence, and this shielding is generally adequate to attenuate the neutrons. However, these neutrons are scattered through the treatment room maze and may result in a radiological problem at the door entrance, a high occupancy area in a radiotherapy facility. In this article, we used MCNP Monte Carlo simulation to calculate neutron doses in the maze of radiotherapy rooms and we suggest an alternative method to the Kersey semi-empirical model of neutron dose calculation at the entrance of mazes. It was found that this new method fits better measured values found in literature, as well as our Monte Carlo simulated ones.
目前,钴和铯远距离治疗机正被直线加速器所取代。这些机器中的最大光子能量可在4至25兆电子伏特之间变化,而且这些设备的一大优势是它们不包含放射源。高能(E>10兆伏)医用直线加速器相对于低能加速器具有若干物理优势:皮肤剂量更低、射线穿透力更强,且靶体积外组织的散射剂量更小。然而,就患者防护和职业剂量而言,高能光子产生的治疗束中不希望有的中子污染已成为一个额外问题。治疗室的墙壁进行了屏蔽,以减弱初级和次级X射线注量,而且这种屏蔽通常足以减弱中子。然而,这些中子会通过治疗室迷宫散射,可能在迷宫入口处导致放射问题,而迷宫入口是放射治疗设施中人员高度密集的区域。在本文中,我们使用MCNP蒙特卡罗模拟来计算放射治疗室迷宫中的中子剂量,并提出一种替代方法,以取代克尔西在迷宫入口处中子剂量计算的半经验模型。结果发现,这种新方法与文献中发现的测量值以及我们的蒙特卡罗模拟值拟合得更好。