Volatile polydimethylcyclosiloxanes in the ambient laboratory air identified as source of extreme background signals in nanoelectrospray mass spectrometry.

作者信息

Schlosser Andreas, Volkmer-Engert Rudolf

机构信息

Institute of Medical Immunology, Charité, Humboldt University, Hessische Strasse 3-4, 10115 Berlin, Germany.

出版信息

J Mass Spectrom. 2003 May;38(5):523-5. doi: 10.1002/jms.465.

Abstract

Polydimethylcyclosiloxanes, an almost ubiquitous air contaminant, can interfere with nanoelectrospray analysis. The sensitivity of nanoelectrospray to these volatile air contaminants was demonstrated in this study. The intensity of the interfering ion signals caused by these compounds can be decreased by changing the position of the nanoESI needle and almost completely suppressed by applying a flow of pure nitrogen around the needle and the sample cone. The nitrogen flow causes a slight shift in charge distribution, but does not influence the sensitivity for peptide detection.