Karim Khairil Juhanni Binti Abd, Jin Ji-Ye, Takeuchi Toyohide
Department of Chemistry, Faculty of Engineering, Gifu University, 1-1 Yanagido, Gifu 501-1193, Japan.
J Chromatogr A. 2003 May 2;995(1-2):153-60. doi: 10.1016/s0021-9673(03)00523-5.
Inorganic anions and cations in environmental waters were determined by ion chromatography. Stationary and mobile phases were examined for the simultaneous separation of both anions and cations. Cations detection by UV detection requires a mobile phase with a UV absorbing additive, which indirectly visualizes cations as negative peaks. Simultaneous separation of anions and cations were achieved when using an eluent that consists of inorganic acid with weak basic amino acid as additives. It was convenient to separate both anions and cations by coupling anion-exchange and cation-exchange columns in tandem. The order of the separation columns connected affected the elution profiles. When the eluent comprises of multiple anions and a single cation, the anion-exchange column should be connected in the upper stream, whereas when the eluent comprises multiple cations and a single anion, the cation-exchange column should be connected in the upper stream. Use of switching valves also allowed simultaneous separation of anions and cations in a single chromatographic run. In the present work, operating conditions were optimized for the simultaneous separation of anions and cations.
采用离子色谱法测定环境水体中的无机阴离子和阳离子。对固定相和流动相进行了研究,以实现阴离子和阳离子的同时分离。通过紫外检测法检测阳离子需要在流动相中添加具有紫外吸收能力的添加剂,该添加剂可将阳离子间接显示为负峰。当使用由无机酸与弱碱性氨基酸作为添加剂组成的洗脱液时,可实现阴离子和阳离子的同时分离。通过串联连接阴离子交换柱和阳离子交换柱来分离阴离子和阳离子很方便。连接的分离柱顺序会影响洗脱曲线。当洗脱液由多种阴离子和单一阳离子组成时,阴离子交换柱应连接在上游;而当洗脱液由多种阳离子和单一阴离子组成时,阳离子交换柱应连接在上游。使用切换阀也可在一次色谱运行中同时分离阴离子和阳离子。在本研究中,对阴离子和阳离子的同时分离操作条件进行了优化。