创世纪支架:一种用于患有先天性心脏病的婴儿、儿童和成人的新型低轮廓支架。

The Genesis stent: A new low-profile stent for use in infants, children, and adults with congenital heart disease.

作者信息

Forbes Thomas J, Rodriguez-Cruz Edwin, Amin Zahid, Benson Lee N, Fagan Thomas E, Hellenbrand William E, Latson Larry A, Moore Phillip, Mullins Charles E, Vincent Julie A

机构信息

Division of Cardiology, Wayne State University/Children's Hospital of Michigan, Detroit, Michigan,USA.

出版信息

Catheter Cardiovasc Interv. 2003 Jul;59(3):406-14. doi: 10.1002/ccd.10547.

Abstract

Placement of intravascular stents that can reach adult size in infants and smaller children has been limited by the large profile and poor flexibility of currently available stents. In vitro and in vivo testing of the Genesis stent was performed to evaluate crimpability, predeployment flexibility, and radial strength. Comparisons were made to the Palmaz iliac and IntraStent (IS) LD stents. Nine physicians placed 30 Genesis stents in swine pulmonary and systemic arteries to evaluate stent deliverability/crimpability. Two swine were recovered and underwent a second catheterization 8 weeks later, where the stents (n = 8) were reexpanded to maximal size. Angiographic and intravascular ultrasound (IVUS) assessments were performed. In vitro testing revealed the Genesis stent to have superior crimpability, flexibility, and comparable radial strength to the Palmaz iliac stent, and superior crimpability and radial strength and comparable flexibility to the IS LD series. During in vivo testing, the physicians graded the Genesis stent superior to the Palmaz stent regarding crimpability and deliverability, and superior to the IS LD stent in regard to crimpability, and comparable to or superior in deliverability. In the chronic animals, the Genesis stent was expanded up to maximal diameter 8 weeks following implantation. Angiographic and IVUS revealed no fractures no in-stent restenosis. The Genesis stent can be easily delivered through smaller sheaths, which will facilitate their use in infants and smaller children with vascular stenosis.

摘要

目前可用的血管内支架外形较大且柔韧性差,限制了其在婴儿和较小儿童体内达到成人尺寸的应用。对Genesis支架进行了体外和体内测试,以评估其可卷曲性、预部署柔韧性和径向强度。并与Palmaz髂动脉支架和IntraStent(IS)LD支架进行了比较。九名医生将30个Genesis支架植入猪的肺动脉和体动脉中,以评估支架的输送性/可卷曲性。回收了两只猪,并在8周后对其进行了第二次导管插入术,将支架(n = 8)重新扩张至最大尺寸。进行了血管造影和血管内超声(IVUS)评估。体外测试显示,Genesis支架的可卷曲性、柔韧性优于Palmaz髂动脉支架,径向强度与之相当;其可卷曲性和径向强度优于IS LD系列,柔韧性与之相当。在体内测试中,医生们认为Genesis支架在可卷曲性和输送性方面优于Palmaz支架,在可卷曲性方面优于IS LD支架,在输送性方面与之相当或更优。在长期观察的动物中,Genesis支架在植入8周后可扩张至最大直径。血管造影和IVUS显示无骨折和支架内再狭窄。Genesis支架可轻松通过较小的鞘管输送,这将便于其用于患有血管狭窄的婴儿和较小儿童。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验