Petrone Matthew R, Guinn Jennifer, Reddin Amanda, Sutlive Thomas G, Flynn Timothy W, Garber Matthew P
US Army, Baylor University Graduate Program in Physical Therapy, USA.
J Orthop Sports Phys Ther. 2003 Jun;33(6):319-25. doi: 10.2519/jospt.2003.33.6.319.
Test-retest reliability and validity.
To determine the validity and reliability of the Palpation Meter (PALM).
Leg length discrepancy (LLD) has been associated with a variety of musculoskeletal disorders. Therefore, the clinical measurement of LLD has become a routine and important part of the physical examination. The PALM is an instrument that was recently developed to indirectly measure LLD, but little is known about its measurement properties.
Fifteen healthy and 15 symptomatic subjects with suspected LLD participated in this study. Measurements of pelvic crest height difference (PD) were obtained by 2 examiners using the PALM. A standing antero-posterior (AP) radiograph of each subject's pelvis was taken, and PD and LLD (femoral head height difference) were determined from the radiograph for comparison with the PALM values. Intraclass correlation coefficients (ICCs) were calculated to determine the validity and reliability estimates of the PALM.
For all subjects, the validity estimates (ICC2,3) of the PALM for PD were excellent (0.90 for rater 1 and 0.92 for rater 2) when compared with the standing AP radiograph of the pelvis. The PALM was less accurate (ICC2,3 of 0.76 and 0.78 for rater 1 and 2, respectively) as an indirect estimate of LLD. Intrarater reliability for each rater was excellent (ICC3,3 = 0.97 and 0.98) and interrater reliability was very good (ICC2,3 = 0.88).
The PALM is a reliable and valid instrument for measuring PD. Clinicians should consider this convenient, cost-effective clinical tool as an alternative to radiographic measurement of pelvic crest height difference.
重测信度和效度。
确定触诊仪(PALM)的效度和信度。
腿长差异(LLD)与多种肌肉骨骼疾病相关。因此,LLD的临床测量已成为体格检查的常规且重要组成部分。PALM是一种最近开发的用于间接测量LLD的仪器,但对其测量特性知之甚少。
15名健康且15名有疑似LLD症状的受试者参与了本研究。两名检查者使用PALM测量骨盆嵴高度差异(PD)。对每个受试者的骨盆进行站立前后位(AP)X线摄影,并从X线片确定PD和LLD(股骨头高度差异),以与PALM值进行比较。计算组内相关系数(ICC)以确定PALM的效度和信度估计值。
对于所有受试者,与骨盆站立AP位X线片相比,PALM对PD的效度估计值(ICC2,3)极佳(检查者1为0.90,检查者2为0.92)。作为LLD的间接估计,PALM的准确性较低(检查者1和2的ICC2,3分别为0.76和0.78)。每位检查者的组内信度极佳(ICC3,3 = 0.97和0.98),检查者间信度非常好(ICC2,3 = 0.88)。
PALM是一种测量PD的可靠且有效的仪器。临床医生应将这种方便、经济高效的临床工具视为骨盆嵴高度差异X线测量的替代方法。