Fahraeus T, Höijer C J
Pacemaker Department, Heart- and Lung Division, University Hospital of Lund, Lund, Sweden.
Europace. 2003 Jul;5(3):279-81. doi: 10.1016/s1099-5129(03)00032-1.
Twiddler's syndrome is a well-known complication of pacemaker treatment. It was first described by Bayliss et al. when a patient manipulated and rotated the pulse generator in the pocket so many turns that it resulted in lead dislodgment, diaphragmatic stimulation and loss of capture. In this case report we present a patient who managed to rotate her dual chamber pulse generator so quickly after implantation that exit block occurred within 17 h. She had wound the two leads as far as their tips in a perfect formation around the pulse generator.
旋弄综合征是起搏器治疗中一种广为人知的并发症。它最早由贝利斯等人描述,当时一名患者在皮下囊袋中对脉冲发生器进行了多次旋转操作,导致导线脱位、膈肌刺激和起搏失效。在本病例报告中,我们介绍了一名患者,其在植入双腔脉冲发生器后很快就将其旋转,以至于在17小时内就出现了出口阻滞。她将两根导线的尖端以完美的形态缠绕在了脉冲发生器周围。