Silvestri Maximilian A, Miles Deborah, Rothwell Arlene P, Wood Karl V, Cushman Mark
Department of Medicinal Chemistry and Molecular Pharmacology, Purdue University, West Lafayette, IN 47907, USA.
Rapid Commun Mass Spectrom. 2003;17(15):1703-8. doi: 10.1002/rcm.1104.
Electrospray ionization (ESI) mass spectrometry was employed to obtain both molecular weight confirmation and structural information for a series of novel alkenyldiarylmethane (ADAM) analogs. The mass spectral data were intended for use during the structure elucidation of ester hydrolysis products formed during an in vitro metabolism study of a series of novel ADAM analogs. The data on the precursor molecules show the presence of the molecular ion peak, M+H, as well as a peak consistent with the hydrolysis product of the original ester (MH-ROH+H(2)O). However, chemical ionization mass spectrometry, elemental analysis and (1)H NMR data indicated the presence of only the intact diester compounds, suggesting that the formation of the hydrolysis product was an instrumental artifact, i.e., in-beam hydrolysis during ESI or a result of longer ion residence times of the ion trap mass analyzer.
采用电喷雾电离(ESI)质谱法来确定一系列新型烯基二芳基甲烷(ADAM)类似物的分子量并获取其结构信息。质谱数据旨在用于一系列新型ADAM类似物体外代谢研究过程中形成的酯水解产物的结构解析。前体分子的数据显示存在分子离子峰M + H,以及与原始酯的水解产物一致的峰(MH - ROH + H₂O)。然而,化学电离质谱法、元素分析和¹H NMR数据表明仅存在完整的二酯化合物,这表明水解产物的形成是一种仪器假象,即在ESI过程中的束内水解或离子阱质谱分析仪中离子停留时间较长的结果。