Suppr超能文献

用于治疗急性和亚急性主动脉下动脉闭塞的旋转血栓切除术装置的早期经验。

Early experience with a rotational thrombectomy device for treatment of acute and subacute infra-aortic arterial occlusions.

作者信息

Zeller Thomas, Frank Ulrich, Bürgelin Karlheinz, Müller Christian, Flügel Peter, Horn Barbara, Schwarzwälder Uwe, Neumann Franz-Josef

机构信息

Department of Angiology, Heart-Center Bad Krozingen, Germany.

出版信息

J Endovasc Ther. 2003 Apr;10(2):322-31. doi: 10.1177/152660280301000224.

Abstract

PURPOSE

To evaluate a rotational thrombectomy device in the treatment of acute and subacute/chronic thrombotic infra-aortic occlusions of native vessels and bypass grafts.

METHODS

From July 2000 to February 2002, 98 patients (65 men; mean age 66+/-9 years, range 47-90) with 100 thrombotic occlusions (mean age of occlusion 31+/-33 days, range 0-140) measuring an average of 21+/-11 cm long (range 2-40) were treated with rotational thrombectomy (Rotarex). There were 33 acute (</=14 days) thrombotic/embolic native artery occlusions (group I), 58 subacute/chronic (>14 days) native artery occlusions (group II), and 9 acute bypass graft occlusions (group III).

RESULTS

The device activation time was 4.9+/-1.4 minutes, during which 4.0+/-1.4 passes of the device were performed. The amount of aspirated fluid was 240+/-119 mL. Slightly less than half the arteries (48%) were stented. Primary success (residual stenosis <30%) was achieved in 92% (94% for group I, 93% for group II, and 78% for group III; 100% for the ipsilateral approach, 56% for the crossover approach). Among the 18 complications, 3 were serious (2 amputations after unsuccessful intervention and 1 death); there were 8 vessel perforations and 7 cases of peripheral embolization. Thirty-day survival and limb salvage was 88% for group I, 100% for group II, and 66% for group III.

CONCLUSIONS

The device is an easy-to-handle, useful tool for ipsilateral treatment of acute and subacute thrombotic arterial and bypass graft occlusions. The use of this device is limited by the 8-F diameter of the catheter and the limited capacity for crossover interventions.

摘要

目的

评估一种旋转血栓切除术装置在治疗天然血管和旁路移植物的急性及亚急性/慢性血栓形成性主动脉下闭塞中的应用。

方法

2000年7月至2002年2月,对98例患者(65例男性;平均年龄66±9岁,范围47 - 90岁)的100处血栓性闭塞病变(闭塞平均年龄31±33天,范围0 - 140天)进行了旋转血栓切除术(Rotarex)治疗,这些病变平均长度为21±11 cm(范围2 - 40 cm)。其中有33例急性(≤14天)血栓形成/栓塞性天然动脉闭塞(I组),58例亚急性/慢性(>14天)天然动脉闭塞(II组),以及9例急性旁路移植物闭塞(III组)。

结果

装置激活时间为4.9±1.4分钟,在此期间装置进行了4.0±1.4次操作。吸出液体量为240±119 mL。略少于一半的动脉(48%)进行了支架置入。主要成功率(残余狭窄<30%)为92%(I组为94%,II组为93%,III组为78%;同侧入路为100%,交叉入路为56%)。在18例并发症中,3例为严重并发症(2例干预失败后截肢和1例死亡);有8例血管穿孔和7例周围栓塞。I组30天生存率和肢体挽救率为88%,II组为100%,III组为66%。

结论

该装置是用于同侧治疗急性和亚急性血栓形成性动脉及旁路移植物闭塞的一种易于操作且有用的工具。该装置的使用受到导管8F直径以及交叉干预能力有限的限制。