van der Zee P, Cope M, Arridge S R, Essenpreis M, Potter L A, Edwards A D, Wyatt J S, McCormick D C, Roth S C, Reynolds E O
Department of Medical Physics, University College London.
Adv Exp Med Biol. 1992;316:143-53. doi: 10.1007/978-1-4615-3404-4_17.
The Differential Pathlength Factor (DPF) has been measured for several different tissues. The results showed that the DPF varied with the type of tissue studied, and in the case of the adult calf with sex. However, the DPF for all tissues studied was constant once the inter optode spacing exceeded 2.5 cm. Thus, measurements can be made by NIR spectroscopy at a range of inter optode spacings, and a single DPF used in the calculation of chromophore concentration. The results also showed that the major source of error in the DPF lay in the measurement of the inter optode spacing. To improve accuracy, two options are possible. Firstly, some means of continuous measurement of inter optode spacing could be incorporated in the NIR instrumentation. The better alternative would be an instrument incorporating a method of directly measuring the optical pathlength at each wavelength. This could be done either by time of flight measurement, or if it can be validated, by phase shift measurement.
已对几种不同组织测量了微分光程因子(DPF)。结果表明,DPF随所研究的组织类型而变化,在成年小牛的情况下还随性别而变化。然而,一旦光极间距超过2.5厘米,所有研究组织的DPF就保持恒定。因此,可以在一系列光极间距下通过近红外光谱法进行测量,并在计算发色团浓度时使用单一的DPF。结果还表明,DPF误差的主要来源在于光极间距的测量。为提高准确性,有两种可行的选择。首先,可以在近红外仪器中加入某种连续测量光极间距的方法。更好的选择是一种采用直接测量每个波长光程长度方法的仪器。这可以通过飞行时间测量来实现,或者如果能够得到验证,也可以通过相移测量来实现。