J Chromatogr A. 2003 Jun 20;1002(1-2):101-9. doi: 10.1016/s0021-9673(03)00729-5.
Size-exclusion chromatography (SEC) combined with online multi-angle light scattering (MALS) and refractometric (RI) detection has been employed for the molar mass characterisation of water-insoluble cationic methyl methacrylate-ethyl acrylate copolymers (Eudragit RS and RL). Due to their positive charge, cationic polymers are particularly difficult to separate on a SEC column, in worst cases being completely adsorbed on the oppositely charged packing material. This work has examined how a careful addition of salt (LiCl) to the copolymer solution in ethanol decreases the electrostatic interactions, clearly seen as a decrease in elution volume from the SEC column as well as an improved recovery. At a certain level of ionic strength, typically about 50 mM, the copolymer recovery from the SEC column reached 100% and molar mass distributions corresponding to the complete sample could be obtained. The combined MALS/RI detection gives the opportunity to measure the absolute molar mass independent of recovery and retention. Thus, in this study, it turned out to be a favourable tool for tracing the changes in elution behaviour of the charged copolymer as the ionic strength was increased.
尺寸排阻色谱法(SEC)结合在线多角度光散射(MALS)和示差折光(RI)检测已用于水不溶性阳离子甲基丙烯酸甲酯-丙烯酸乙酯共聚物(尤特奇RS和RL)的摩尔质量表征。由于阳离子聚合物带正电荷,在SEC柱上特别难以分离,在最坏的情况下会完全吸附在带相反电荷的填充材料上。这项工作研究了如何在乙醇中的共聚物溶液中小心添加盐(LiCl)来减少静电相互作用,这明显表现为SEC柱洗脱体积的减少以及回收率的提高。在一定的离子强度水平下,通常约为50 mM,从SEC柱上的共聚物回收率达到100%,并且可以获得对应于完整样品的摩尔质量分布。MALS/RI联合检测提供了独立于回收率和保留率测量绝对摩尔质量的机会。因此,在本研究中,它被证明是追踪随着离子强度增加带电共聚物洗脱行为变化的有利工具。