Gritti Fabrice, Guiochon Georges
Department of Chemistry, The University of Tennessee, Knoxville, TN 37996-1600, USA.
J Chromatogr A. 2003 Aug 1;1008(1):13-21. doi: 10.1016/s0021-9673(03)00972-5.
A series of samples of increasing volume (from 0.001 to 4.0 cm3), containing the same constant concentration (40 g/l) of two simple compounds, ethylbenzoate and 4-tert.-butylphenol were injected on a Kromasil-C18 column with methanol-water (62:38. v/v) as the mobile phase. Complex band profiles were observed when the volume of the sample became large enough and strong band interference took place. The analysis of the fractions collected during the elution of the mixed band demonstrates that, for samples larger than 2 cm3, the band of 4-tert.-butylphenol is split into two separate bands, one eluted before and the other eluted after the band of ethylbenzoate. Such a phenomenon has never been observed yet in RPLC, under isocratic elution conditions.