Duan Jun, Shen Sui, Fiveash John B, Brezovich Ivan A, Popple Richard A, Pareek Prem N
Department of Radiation Oncology, University of Alabama Birmingham, 619 South 19th Street, Birmingham, Alabama 35233, USA.
Med Phys. 2003 Aug;30(8):2241-52. doi: 10.1118/1.1592017.
Intensity modulated radiation therapy (IMRT) with a dynamic multileaf collimator (DMLC) requires synchronization of DMLC leaf motion with dose delivery. A delay in DMLC communication is known to cause leaf lag and lead to dosimetric errors. The errors may be exacerbated by gated operation. The purpose of this study was to investigate the effect of leaf lag on the accuracy of doses delivered in gated IMRT. We first determined the effective leaf delay time by measuring the dose in a stationary phantom delivered by wedge-shaped fields. The wedge fields were generated by a DMLC at various dose rates. The so determined delay varied from 88.3 to 90.5 ms. The dosimetric effect of this delay on gated IMRT was studied by delivering wedge-shaped and clinical IMRT fields to moving and stationary phantoms at dose rates ranging from 100 to 600 MU/min, with and without gating. Respiratory motion was simulated by a linear sinusoidal motion of the phantom. An ionization chamber and films were employed for absolute dose and 2-D dose distribution measurements. Discrepancies between gated and nongated delivery to the stationary phantom were observed in both absolute dose and 2-D dose distribution measurements. These discrepancies increased monotonically with dose rate and frequency of beam interruptions, and could reach 3.7% of the total dose delivered to a 0.6 cm3 ion chamber. Isodose lines could be shifted by as much as 3 mm. The results are consistent with the explanation that beam hold-offs in gated delivery allowed the lagging leaves to catch up with the delivered monitor units each time that the beam was interrupted. Low dose rates, slow leaf speeds and low frequencies of beam interruptions reduce the effect of this delay-and-catch-up cycle. For gated IMRT it is therefore important to find a good balance between the conflicting requirements of rapid dose delivery and delivery accuracy.
使用动态多叶准直器(DMLC)的调强放射治疗(IMRT)需要DMLC叶片运动与剂量输送同步。已知DMLC通信延迟会导致叶片滞后并导致剂量学误差。门控操作可能会使这些误差加剧。本研究的目的是调查叶片滞后对门控IMRT中输送剂量准确性的影响。我们首先通过测量楔形野在静止模体中输送的剂量来确定有效叶片延迟时间。楔形野由DMLC在不同剂量率下生成。如此确定的延迟在88.3至90.5毫秒之间变化。通过在100至600 MU/分钟的剂量率下,对运动和静止模体输送楔形野和临床IMRT野,研究了这种延迟对门控IMRT的剂量学影响,包括有门控和无门控情况。通过模体的线性正弦运动模拟呼吸运动。使用电离室和胶片进行绝对剂量和二维剂量分布测量。在绝对剂量和二维剂量分布测量中均观察到门控和非门控输送到静止模体之间的差异。这些差异随剂量率和束流中断频率单调增加,并且可能达到输送到0.6 cm³电离室的总剂量的3.7%。等剂量线可能会偏移多达3毫米。结果与以下解释一致:门控输送中的束流暂停使得每次束流中断时滞后的叶片能够赶上已输送的监测单位。低剂量率、慢叶片速度和低束流中断频率可降低这种延迟和追赶循环的影响。因此,对于门控IMRT,在快速剂量输送和输送准确性这两个相互冲突的要求之间找到良好平衡很重要。