Suppr超能文献

确定用于最小化人类额下回皮质易感性伪影的最佳口腔内被动垫片的方案。

Protocol to determine the optimal intraoral passive shim for minimisation of susceptibility artifact in human inferior frontal cortex.

作者信息

Wilson James L, Jenkinson Mark, Jezzard Peter

机构信息

Centre for Functional Magnetic Resonance Imaging of the Brain, Department of Clinical Neurology, University of Oxford, John Radcliffe Hospital, Oxford OX3 9DU, UK.

出版信息

Neuroimage. 2003 Aug;19(4):1802-11. doi: 10.1016/s1053-8119(03)00225-8.

Abstract

At higher static magnetic field (B(0)) strengths (>/=3 T), the study of human inferior frontal cortex (IFC) when utilising a variety of MRI techniques is severely hampered by the presence of susceptibility artifacts. This is particularly the case for blood oxygenation level-dependent functional MRI, where large signal voids are generally encountered in the frontal lobes. A previous study described an approach to artifact correction involving a mouth insert consisting of a prototype diamagnetic passive shim [Magn. Reson. Med. 48 (2002), 906]. Here we extend that method by investigating the effect of five different intraoral passive shims on B(0) homogeneity and echoplanar imaging susceptibility artifacts within the brain, and particularly the IFC, of six subjects. The optimal passive shim is shown to be subject- and study-specific, providing an average reduction in mean absolute B(0) offset within the IFC of 57%, along with a concomitant reduction in echoplanar susceptibility artifact. All subjects were at ease while wearing the intraoral shims. A 4-min in vivo protocol to determine the optimal passive shim from the available set, utilising intrinsic structural and B(0) subject data, is described and shown to be accurate and reliable.

摘要

在较高的静磁场(B(0))强度(≥3 T)下,利用各种磁共振成像(MRI)技术对人类额下回(IFC)进行研究时,由于存在磁化率伪影而受到严重阻碍。对于血氧水平依赖性功能MRI而言尤其如此,在额叶通常会遇到大的信号空洞。先前的一项研究描述了一种伪影校正方法,该方法涉及一种由原型抗磁性无源匀场片组成的口腔插入物[《磁共振医学》48(2002),906]。在此,我们通过研究五种不同的口腔内无源匀场片对六名受试者大脑内,尤其是IFC内的B(0)均匀性和回波平面成像磁化率伪影的影响,对该方法进行了扩展。结果表明,最佳的无源匀场片因受试者和研究而异,可使IFC内的平均绝对B(0)偏移平均降低57%,同时回波平面磁化率伪影也相应减少。所有受试者在佩戴口腔内匀场片时都感觉舒适。本文描述了一种4分钟的体内方案,该方案利用内在结构和B(0)受试者数据从可用的匀场片中确定最佳无源匀场片,结果表明该方案准确可靠。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验