James J M, Lloyd J J, Leahy B C, Shields R A, Prescott M C, Testa H J
Department of Nuclear Medicine and Medical Physics, Manchester Royal Infirmary, UK.
Br J Radiol. 1992 May;65(773):403-8. doi: 10.1259/0007-1285-65-773-403.
Inhalation of the ventilatory radiopharmaceutical 99Tcm Technegas leads, in some patients, to symptoms that may be attributed to temporary lowering of oxygen saturation. In order to evaluate this, oxygen saturation was measured by pulse oximetry in a series of patients undergoing Technegas ventilation scintigraphy. A decrease in oxygen saturation was recorded in 87% of the patient group. The mean change, as a percentage of the initial value, was 8.3% (range 1-24%). Hypoxia arising in association with Technegas administration may be reduced by pre-oxygenation. In patients who were pre-oxygenated, oxygen saturation did not fall below 85% (PaO2: 50 mmHg) but in 39% of those not pre-oxygenated the value fell below this level. Oxygen saturation was also monitored in a series of patients undergoing perfusion scintigraphy. In 17% of patients a decrease was recorded (range 2-11%). In view of the large number of perfusion scans performed annually in this department and elsewhere without untoward effect, such temporary decreases in oxygen saturation presumably present no hazard to the patient.
吸入通气放射性药物99Tcm锝气体,在一些患者中会引发一些症状,这些症状可能归因于氧饱和度的暂时降低。为了对此进行评估,在一系列接受锝气体通气闪烁扫描的患者中,通过脉搏血氧饱和度仪测量了氧饱和度。在患者组中,87%的患者记录到氧饱和度下降。平均变化量,以初始值的百分比计,为8.3%(范围1 - 24%)。与锝气体给药相关的缺氧情况可通过预给氧来减轻。在预给氧的患者中,氧饱和度未降至85%以下(动脉血氧分压:50 mmHg),但在未预给氧的患者中,39%的患者该值降至该水平以下。在一系列接受灌注闪烁扫描的患者中也监测了氧饱和度。17%的患者记录到氧饱和度下降(范围2 - 11%)。鉴于该科室及其他地方每年进行大量灌注扫描且未产生不良影响,这种氧饱和度的暂时下降大概对患者没有危害。