Suppr超能文献

短时间高压暴露期间氙从兔股骨中的清除情况。

Xenon washout from the rabbit femur during short hyperbaric exposures.

作者信息

Davis T R

机构信息

Queens Medical Centre, Nottingham University, England.

出版信息

Undersea Biomed Res. 1992 Sep;19(5):355-9.

PMID:1355313
Abstract

133Xenon washout from the femora of 5 anesthetized rabbits was recorded during short hyperbaric exposures (3 atm abs). Equipment tests showed that the scintillation counter was heat sensitive. The recorded count rate from a constant source of 133xenon decreased during compression (temperature rose 5 degrees C) and increased during decompression (temperature fell 5 degrees C). When the scintillation counter was thermally insulated, the rate of xenon washout from the femur remained unchanged in all rabbits during these hyperbaric exposures. The conclusion is that the rate of xenon washout from the femur is not affected by changes in ambient pressure. As most scintillation counters are heat sensitive, it is possible that the previous report of such changes was erroneous and caused by heat sensitivity of the recording equipment.

摘要

在短时间高压暴露(3个绝对大气压)期间,记录了5只麻醉兔股骨中133氙的洗脱情况。设备测试表明,闪烁计数器对热敏感。在压缩过程中(温度上升5摄氏度),来自恒定133氙源的记录计数率下降,而在减压过程中(温度下降5摄氏度)则上升。当闪烁计数器进行热绝缘时,在这些高压暴露期间,所有兔子股骨中氙的洗脱速率保持不变。结论是,股骨中氙的洗脱速率不受环境压力变化的影响。由于大多数闪烁计数器对热敏感,之前关于此类变化的报告有可能是错误的,是由记录设备的热敏感性导致的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验