Suppr超能文献

基于单克隆抗体的酶联免疫吸附测定法(ELISA)用于测定维生素K依赖性蛋白S:抗体免疫反应性对血浆蛋白S抗原测定的影响。

Monoclonal antibody-based enzyme-linked immunosorbent assays (ELISA) for the measurement of vitamin K-dependent protein S: the effect of antibody immunoreactivity on plasma protein S antigen determinations.

作者信息

D'Angelo S V, Tombesi S, Marcovina S, Albertini A, Della Valle P, D'Angelo A

机构信息

Coagulation Service, Scientific Institute IRCCS H. S. Raffaele, Milano, Italy.

出版信息

Thromb Haemost. 1992 Jun 1;67(6):631-8.

PMID:1387263
Abstract

Two monoclonal antibodies (Mabs) specifically directed to human protein S (PS) - named 5E9E9 and 3B10.25 - were produced and their properties compared to those of 2 previously characterized anti-PS-Mabs (HPS-2 and S10). 3B10.25, similar to S10, was directed to the calcium-free conformation of PS and had virtually identical affinity for free and C4b-binding protein (C4b-BP)-bound PS; 5E9E9 similar to HPS-2, had no calcium-dependency and was selectively directed to free PS. All Mabs were equally reactive to freshly purified and thrombin-cleaved PS. To evaluate the influence of C4b-BP bound PS on PS antigen determinations, ELISA systems employing the four Mabs individually as capture antibody (Ab) and peroxidase-conjugated polyclonal anti-PS IgG as detecting Ab were developed and compared to immunoelectrophoresis (EIA) and to an ELISA employing polyclonal anti-PS IgG as capture and detecting Ab, in the determination of PS in purified systems and in plasma. With all the ELISAs there was parallelism of dilution curves obtained with normal plasma and purified PS; however, supplementation of plasma with purified C4b-BP resulted in loss of parallelism when employing the Mabs directed to free PS as capture Ab. Influence of high C4b-BP on PS antigen determinations was confirmed in a series of plasma samples from patients with C4b-BP levels ranging from 70% to over 200%. Compared to the values obtained with the S10- or 3B10.25 - based ELISAs - which were similar despite a 10-fold difference in sample dilution - plasma PS was underestimated by the ELISAs employing 5E9E9 or HPS-2 while it was overestimated by EIA.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

制备了两种特异性针对人蛋白S(PS)的单克隆抗体(Mab)——命名为5E9E9和3B10.25,并将它们的特性与之前鉴定的两种抗PS单克隆抗体(HPS-2和S10)进行比较。3B10.25与S10相似,针对PS的无钙构象,对游离PS和与C4b结合蛋白(C4b-BP)结合的PS具有几乎相同的亲和力;5E9E9与HPS-2相似,无钙依赖性,且选择性地针对游离PS。所有单克隆抗体对新鲜纯化的和经凝血酶切割的PS的反应性相同。为了评估与C4b-BP结合的PS对PS抗原测定的影响,开发了分别以这四种单克隆抗体作为捕获抗体(Ab)、过氧化物酶偶联的多克隆抗PS IgG作为检测抗体的ELISA系统,并在纯化系统和血浆中测定PS时,与免疫电泳(EIA)以及以多克隆抗PS IgG作为捕获和检测抗体的ELISA进行比较。在所有ELISA中,用正常血浆和纯化PS获得的稀释曲线具有平行性;然而,当使用针对游离PS的单克隆抗体作为捕获抗体时,向血浆中添加纯化的C4b-BP会导致平行性丧失。在一系列C4b-BP水平从70%到超过200%的患者血浆样本中,证实了高C4b-BP对PS抗原测定的影响。与基于S10或3B10.25的ELISA所获得的值相比——尽管样品稀释相差10倍,但二者相似——使用5E9E9或HPS-2的ELISA低估了血浆PS,而EIA高估了血浆PS。(摘要截短于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验