Suppr超能文献

微孔板凝血试验。

Microplate coagulation assays.

作者信息

Pratt C W, Monroe D M

机构信息

Division of Hematology, University of North Carolina, Chapel Hill 27599-7035.

出版信息

Biotechniques. 1992 Sep;13(3):430-3.

PMID:1389175
Abstract

This report describes the development of microplate-based blood coagulation assays. The assays require a kinetic microplate reader to follow changes in absorbance at 405 nm caused by the coagulating plasma. Procedures for performing prothrombin time and activated partial thromboplastin time tests are described with intra- and inter-assay variability of a few percentage points. The prothrombin time of normal plasma was 64.5 +/- 3.6 s, and the activated partial thromboplastin time was 69.8 +/- 3.2 s. Clotting times were prolonged when normal plasma was mixed with plasmas deficient in particular coagulation factors, as expected. These assays take advantage of the microplate format (small sample size and multiple simultaneous assays) and can be customized for specific purposes, such as quantifying purified factor IX or assessing protein C activity in plasma.

摘要

本报告描述了基于微孔板的血液凝固检测方法的开发。这些检测需要一台动力学微孔板读数仪来跟踪由凝固血浆引起的405nm处吸光度的变化。文中描述了进行凝血酶原时间和活化部分凝血活酶时间检测的程序,批内和批间变异率为几个百分点。正常血浆的凝血酶原时间为64.5±3.6秒,活化部分凝血活酶时间为69.8±3.2秒。正如预期的那样,当正常血浆与特定凝血因子缺乏的血浆混合时,凝血时间会延长。这些检测利用了微孔板形式(样本量小且可同时进行多个检测)的优势,并且可以针对特定目的进行定制,例如定量纯化的因子IX或评估血浆中的蛋白C活性。