Effects of niobium filtration and constant potential on image quality in dental radiography. 2. Objective assessment.

作者信息

Price C, McDonnell D

机构信息

University of British Columbia, Vancouver.

出版信息

Dentomaxillofac Radiol. 1992 Feb;21(1):5-8. doi: 10.1259/dmfr.21.1.1397453.

Abstract

The effects of niobium filtration and constant potential on image quality were explored by asking observers to identify numbers of circular areas of small density differences in zones of high and low density. Successful identification was related negatively to half-value layer. This relationship was particularly strong in the case of self-rectified units without additional niobium filtration and the constant potential units (r2 = 0.994, P < 0.0005). Constant potential had no effect other than in relation to half-value layer. Additional niobium filtration detracted from successful identification, but slightly less than would be predicted by the half-value layer. Whether this difference is significant or relevant has not been established. The findings of this study, in conjunction with our previous investigations, have failed to demonstrate any clear indications for preferring constant potential supply over self-rectification or for using niobium as an additional filter material. Half-value layer has been shown to be a reliable indicator of the effects of beam quality on emulsion speed, attenuation in water, and image quality. It appears that the methods employed to change the beam quality have no specific effects.

摘要

相似文献

3
Effects of additional filtration on image quality in dental radiography: comparison of niobium with aluminium.
Dentomaxillofac Radiol. 1992 May;21(2):73-6. doi: 10.1259/dmfr.21.2.1397460.
4
Effects of niobium filtration and constant potential on the attenuation of dental X-ray beams by water.
Dentomaxillofac Radiol. 1991 May;20(2):77-80. doi: 10.1259/dmfr.20.2.1936421.
5
Effects of niobium filtration and constant potential on the sensitometric responses of dental radiographic films.
Dentomaxillofac Radiol. 1991 Feb;20(1):11-6. doi: 10.1259/dmfr.20.1.1884846.
7
Diagnostic image quality and dose reduction using niobium filtration for cephalometric radiography.
Dentomaxillofac Radiol. 1993 Nov;22(4):189-94. doi: 10.1259/dmfr.22.4.8181645.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验