Shi Z, Boccalon H, Elias A, Garcia-Serrano A
CHU Rangueil, Toulouse, France.
Int Angiol. 1992 Apr-Jun;11(2):127-31.
A new strain-gauge plethysmographic system, Phlebotest, was evaluated in 19 normal subjects, yielding 38 lower limbs for study. A capacitance mode was used which allowed both the left and right calves to reach their maximal volume expansion before releasing the cuff pressure (60 mmHg in this study). Parameters such as the maximum volume change (V sec), the outflow rate during the first second after deflation (F 1.0), the expelled volume in four seconds (EV 4.0) and the surface area over the curve during the first four seconds after deflation (IND) were automatically calculated and their side differences between the left and right legs were determined respectively. There was a good linear correlation between EV 4.0 and V sec (r = 0.9274, p less than 0.0005) and the EV 4.0/V sec ratio was calculated. In contrast to the previous reports, between three consecutive determinations there were no significant differences in the measurements of these parameters, and V sec consistently kept an identical mean value. These results suggest that using this new system only one determination of calf expansion is required. The reason might be that, in addition to some technical improvements, the computer-controlled capacitance function can secure the optimal venous filling in every determination.
一种新型应变片体积描记系统——静脉检测系统(Phlebotest),在19名正常受试者身上进行了评估,共获得38条下肢用于研究。采用电容模式,在释放袖带压力(本研究中为60 mmHg)之前,使左右小腿均达到最大体积扩张。自动计算诸如最大体积变化(V sec)、放气后第一秒的流出率(F 1.0)、四秒内排出的体积(EV 4.0)以及放气后前四秒曲线的表面积(IND)等参数,并分别确定左右腿之间的差异。EV 4.0与V sec之间存在良好的线性相关性(r = 0.9274,p小于0.0005),并计算了EV 4.0/V sec比值。与先前的报告不同,在连续三次测定中,这些参数的测量值没有显著差异,且V sec始终保持相同的平均值。这些结果表明,使用这种新系统只需测定一次小腿扩张情况。原因可能是,除了一些技术改进外,计算机控制的电容功能能够确保每次测定中静脉充盈达到最佳状态。