Reliability of skinfold calipers as a tool for measuring body fat in human beings.

作者信息

Walia B N, Bhalla A K, Suri S

机构信息

Department of Paediatrics, Postgraduate Institute of Medical Education & Research, Chandigarh.

出版信息

Indian J Med Res. 1992 Aug;96:255-7.

PMID:1428065
Abstract

Triceps skinfold thickness measured in ten adults aged 19 to 50 yr by a Holtain skinfold caliper, was compared with subcutaneous fat thickness obtained on cross-sections of the left upper arm by computerized axial tomography (CAT). The thickness of subcutaneous fat was measured on CAT scan film directly with the inbuilt computer graphic technique. The measurements taken on CAT scan film were found to be significantly higher (by 21-45%; P less than 0.001) than those by the skinfold calipers. Our findings suggest that skinfold calipers may not be very accurate tools for the assessment of body fat in human beings. Therefore, its use in the determination of body composition, for which great precision is required calls for re-thinking.