de Vismes A, Amiot M N
BNM-CEA/LNHB, Laboratoire National Henri Becquerel, Gif-sur-Yvette Cedex 91191, France.
Appl Radiat Isot. 2003 Oct;59(4):267-72. doi: 10.1016/s0969-8043(03)00173-8.
The Monte-Carlo code PENELOPE for Ionisation Chamber Simulation Method has been applied for the calculation of ionisation-chamber (IC) calibration factors. Measuring only a few radionuclides well selected within the relevant energy range, and determining an adjustable parameter, commonly used radionuclides can be measured without any specific calibration. The simulation revealed a discontinuity in the IC response as a function of photon energy and its dependence on the chemical composition of the radioactive solution.