Acute and chronic performance of a steroid eluting electrode for ventricular pacing.

作者信息

Rhoden W E, Llewellyn M J, Schofield S W, Bennett D H

机构信息

Regional Cardiac Centre, Wythenshawe Hospital, Manchester, UK.

出版信息

Int J Cardiol. 1992 Nov;37(2):209-12. doi: 10.1016/0167-5273(92)90210-t.

Abstract

Studies have demonstrated consistently low stimulation thresholds in the first months following implantation of steroid eluting electrodes. There have been no reported cases of exit block. There is little information on the long-term performance of these leads. Data were collected on 15 patients in whom Medtronic steroid eluting leads (5023 and 4003) had been implanted in 1987. Stimulation thresholds were measured at implantation and at weeks 1, 2, 3, 4, 5, 6 and 12. Stimulation threshold, lead impedance and sensing thresholds were monitored periodically for 3 yr using the telemetered function of the pulse generator. Three-month follow-up was obtained in all the patients and 3-yr follow-up in 10. Stimulation thresholds remained low throughout the follow-up period in all but one patient who developed exit block necessitating the replacement of the pacing system. In the remaining patients there was a significant increase in mean stimulation threshold from implantation to week 1 (mean stimulation threshold at implantation = 0.106 ms, mean ST at week 1 = 0.182 ms; p = 0.05), and also a gradual increase in the stimulation threshold over the first 2 yr of follow up (mean stimulation threshold 1 week = 0.182 ms, mean stimulation threshold months 13-24 = 0.243 ms; p = 0.05). Lead impedance showed no significant change over the follow-up period (mean change in impedance = 4.3 omega, SD 163.2, p = 0.97). Unlike many conventional electrodes, steroid eluting leads show no early peak in stimulation threshold but exit block can occur. Stimulation thresholds show a small but significant rise over time.

摘要

研究表明,在植入类固醇洗脱电极后的最初几个月里,刺激阈值一直很低。目前尚无传出阻滞的报道病例。关于这些导线的长期性能的信息很少。收集了1987年植入美敦力类固醇洗脱导线(5023和4003)的15名患者的数据。在植入时以及第1、2、3、4、5、6和12周测量刺激阈值。使用脉冲发生器的遥测功能对刺激阈值、导线阻抗和感知阈值进行了3年的定期监测。所有患者均获得了3个月的随访,10名患者获得了3年的随访。除1例发生传出阻滞需要更换起搏系统的患者外,所有患者在整个随访期间刺激阈值均保持较低。在其余患者中,从植入到第1周平均刺激阈值有显著增加(植入时平均刺激阈值=0.106毫秒,第1周平均刺激阈值=0.182毫秒;p=0.05),并且在随访的前2年中刺激阈值也逐渐增加(第1周平均刺激阈值=0.182毫秒,第13 - 24个月平均刺激阈值=0.243毫秒;p=0.05)。随访期间导线阻抗无显著变化(阻抗平均变化=4.3欧姆,标准差163.2,p=0.97)。与许多传统电极不同,类固醇洗脱导线在刺激阈值上没有早期峰值,但可能会发生传出阻滞。刺激阈值随时间有小幅但显著的上升。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验