Riter Leah S, Meurer Eduardo C, Cotte-Rodriguez Ismael, Eberlin Marcos N, Cooks R Graham
Chemistry Department, Purdue University, West Lafayette, IN 47907, USA.
Analyst. 2003 Sep;128(9):1119-22. doi: 10.1039/b308292j.
Fiber introduction mass spectrometry (FIMS), a variation of solid-phase microextraction (SPME) and membrane introduction mass spectrometry (MIMS), is employed with a miniature mass spectrometer. The inlet system, constructed of commercially available vacuum parts, allows the direct introduction of the SPME needle vacuum chamber into the mass spectrometer. Thermal desorption of the analyte from the poly(dimethylsiloxane) (PDMS) coated fiber was achieved with a built in nichrome heater, followed by electron ionization of the analytes internal to the cylindrical ion trap (CIT). The system has been tested with several volatile organic compounds (VOC) in air and to analyze the headspace over aqueous solutions, with limits of detection in the low ppb range. The signal rise (10-90%) and fall (90-10%) times for the system ranged from 0.1 to 1 s (rise) and 1.2 to 6 s (fall) using heated desorption. In addition, this method has been applied to quantitation of toluene in benzene, toluene, xylene (BTX) mixtures in water and gasoline. This simple and rapid analysis method, coupled to a portable mass spectrometer, has been shown to provide a robust, simple, rapid, reproducible, accurate and sensitive (low ppb range) fieldable approach to the effective in situ analysis of VOC in various matrices.
纤维进样质谱法(FIMS)是固相微萃取(SPME)和膜进样质谱法(MIMS)的一种变体,与微型质谱仪联用。进样系统由市售真空部件构成,可将SPME针真空腔直接引入质谱仪。通过内置的镍铬合金加热器实现分析物从聚二甲基硅氧烷(PDMS)涂层纤维上的热脱附,随后在圆柱形离子阱(CIT)内部对分析物进行电子电离。该系统已针对空气中的几种挥发性有机化合物(VOC)进行了测试,并用于分析水溶液上方的顶空,检测限在低ppb范围内。使用加热脱附时,该系统的信号上升(10 - 90%)和下降(90 - 10%)时间范围为0.1至1秒(上升)和1.2至6秒(下降)。此外,该方法已应用于水和汽油中苯、甲苯、二甲苯(BTX)混合物中甲苯的定量分析。这种简单快速的分析方法与便携式质谱仪联用,已被证明能为各种基质中VOC的有效现场原位分析提供一种稳健、简单、快速、可重现、准确且灵敏(低ppb范围)的方法。