Alnajjar Ahmed, McCord Bruce
Department of Chemistry and Biochemistry, Clippinger Laboratories, Ohio University, Athens, OH 45701, USA.
J Pharm Biomed Anal. 2003 Oct 15;33(3):463-73. doi: 10.1016/s0731-7085(03)00298-x.
A method has been developed for the detection of a mixture of morphine, codeine, 6-acetyl morphine (6-AM) and normorphine using capillary zone electrophoresis (CZE). The method utilized urinary 6-AM as a diagnostic indicator of heroin abuse because it is not a product of either morphine or codeine metabolism. The electrophoretic separation was achieved using an uncoated (50 microm I.D.) fused-silica capillary, 77 cm long, containing the detector window 10.0 cm from the outlet end. The running buffer (pH 6.0) contained 50 mM sodium phosphate and 0.015 M beta-cyclodextrins (beta-CD). The samples were first extracted using a mixed-mode solid-phase extraction procedure and then analyzed by CZE. The UV absorbance detection was monitored at 214 nm. It has been found that beta-CDs can improve separation efficiency due to their hydrophobic cavity. The effect of the concentration of beta-CD and pH was also evaluated. The application of electrokinetic injection with field amplified sample stacking results in low detection limits (40 ng/ml for each analyte) and the method has good reproducibility, precision, accuracy, and high recovery.
已开发出一种利用毛细管区带电泳(CZE)检测吗啡、可待因、6-乙酰吗啡(6-AM)和去甲吗啡混合物的方法。该方法将尿液中的6-AM用作海洛因滥用的诊断指标,因为它不是吗啡或可待因代谢的产物。电泳分离是使用一根未涂层(内径50微米)、长77厘米的熔融石英毛细管实现的,检测窗口位于距出口端10.0厘米处。运行缓冲液(pH 6.0)含有50 mM磷酸钠和0.015 Mβ-环糊精(β-CD)。样品首先采用混合模式固相萃取程序进行萃取,然后通过CZE进行分析。在214 nm处监测紫外吸光度检测。已发现β-CD因其疏水腔可提高分离效率。还评估了β-CD浓度和pH的影响。采用电动进样和场放大样品堆积可实现低检测限(每种分析物为40 ng/ml),该方法具有良好的重现性、精密度、准确度和高回收率。